High Power Electronic Load 63200A Series Operation & Programming Manual
6-4
632xxA-150-xxx
150V
DMM (V)±0.045V
15V
DMM (V)±0.0247V
Checking Medium Voltage Range
A. After tested the high voltage range, press
RANGE
to M range.
B. The DC Source voltage outputs the voltage/current values listed in Table 6-6.
C. Wait for 30 seconds after the DC Source outputted and to record the voltage measured
by DMM (V) and the Load.
Table 6-6
Model
DC Source
Output Voltage
Front Panel Display
Reading
632xxA-150-xxx
80V
DMM (V)±0.024V
8V
DMM (V)±0.0132V
Checking Low Voltage Range
A. After tested the medium voltage range, press
RANGE
to L range.
B. The DC Source voltage outputs the voltage/current values listed in Table 6-7.
C. Wait for 30 seconds after the DC Source outputted and to record the voltage measured
by DMM (V) and the Load.
Table 6-7
Model
DC Source
Output Voltage
Front Panel Display
Reading
632xxA-150-xxx
16V
DMM (V)±0.0048V
1.6V
DMM (V)±0.00264V