73
Senior Solutions™ Therapy System
8 ULTRASOUND CALIBRATION
8.1 GENERAL
A. Tools and Equipment Required
Senior Solutions Combination Therapy
System and all Senior Solutions Ultrasound
Applicators associated with the System being
serviced.
Ohmic Instruments UPM DT 10 or DT 100
Ultrasound Power Meter, set to “watts”.
Degassed Water.
Refer to page 31
for
Degassed Water Recipes.
B. Ultrasound Applicator Calibration
Procedures
Enter the Technical Service screen of the
Therapy System by pressing the STOP,
PAUSE, and START buttons simultaneously.
See Figure 8.1.
NOTE:
To access the Technical Service screen of
Therapy Systems with Version 2.0 or later
software, simultaneously press the HOME
and BACK button from within the Utilities
screen.
See inset at Figure 8.1.
Set up Power Meter per Ohmic User Manual.
Position the Ultrasound Applicator in the
Power Meter.
Press the Ultrasound Calibration button.
See
Figure 8.2.
Press the Head Size button until the size
applicator being calibrated is displayed.
See
Figure 8.3.
Press the START button, refer to
Figure 8.3.
Follow the instructions displayed on the
Therapy System.
Repeat this procedure for each Ultrasound
Applicator associated with the Therapy
System being serviced.
1.
2.
3.
1.
2.
3.
4.
5.
6.
FIGURE 8.1
FIGURE 8.2
FIGURE 8.3
ULTRASOUND CALIBRATION
BUTTON
HEAD SIZE BUTTON
START
BUTTON
THERAPY SYSTEMS WITH
SOFTWARE VERSION 2.0 OR
ABOVE
USE ONLY DEGASSED WATER IN POWER METER FOR
CALIBRATING ULTRASOUND APPLICATORS.
USE OF OTHER TYPES OF WATER WILL CAUSE FALSE
READINGS AND BAD TEST RESULTS.
SEE PAGE 31
FOR DEGASSED WATER RECIPES.
USE OF OTHER BRANDS OR TYPES OF TOOLS,
EQUIPMENT, FIXTURES, MATERIALS, AND SUPPLIES
OTHER THAN THOSE SPECIFICALLY LISTED IN “A. Tools
and Equipment Required” ABOVE WILL GIVE BAD
TEST AND CALIBRATION RESULTS.
IF PROPER EQUIPMENT IS NOT AVAILABLE OR
CAN NOT BE OBTAINED, SEND THE ULTRASOUND
APPLICATORS TO THE FACTORY FOR CALIBRATION.
Содержание SENIOR SOLUTIONS
Страница 1: ...ISO 13485 CERTIFIED Therapy System SERVICEMANUAL Moving Rehabilitation Foward ...
Страница 2: ......
Страница 5: ......
Страница 85: ...80 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONTROL BOARD 27053 Sheet 1 ...
Страница 86: ...81 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONTROL BOARD 27053 Sheet 2 ...
Страница 87: ...82 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONTROL BOARD 27052 Sheet 3 ...
Страница 91: ...86 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 1 ...
Страница 92: ...87 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 2 ...
Страница 93: ...88 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 3 ...
Страница 94: ...89 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 4 ...
Страница 95: ...90 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 5 ...
Страница 96: ...91 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 6 ...
Страница 97: ...92 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 7 ...
Страница 98: ...93 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 8 ...
Страница 99: ...94 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 9 ...
Страница 100: ...95 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 10 ...
Страница 101: ...96 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONNECTOR BOARD 27059 ...
Страница 104: ...99 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM LASER MODULE BOARD 27069 ...
Страница 105: ......
Страница 106: ......