62

 Senior Solutions™ Therapy System

6 REMOVAL/REPLACEMENT

6.6 THERAPY SYSTEM 
 CONTROL 

BOARD 

ASSEMBLY

 

A. Tools and Equipment Required

#1 Phillips Screwdriver
Needle Nose Pliers

 

B. Control Board Assembly Removal

Separate Top from Bottom. 

Refer to 6.4, 

part B.

Remove Fan. 

Refer to 6.5, part B.

Remove the Contrast Knob.

 See Figure 6.25.

Remove the Control Board Ribbon Cable and 
the two Control Board Assembly Retaining 
Screws. 

See Figure 6.26.

STARTing at the top of the Control Board 
Assembly, lift Control Board with one hand, 
push in and release the plastic clips holding 
Control Board Assembly in position. 

See 

Figure 6.27.

While holding up on Control Board Assembly, 
use the Needle Nose Pliers to squeeze the 
lower plastic retainers holding the bottom of 
the Control Board. 

See Figure 6.27

 inset.

Remove the Card Reader Housing from the 
Control Board Assembly.

 

C. Replacing Control Board Assembly

Replace new Control Board Assembly, part 
number 27053, by reversing the 

steps 1- 7 

above.

NOTE:

Do not over tighten the screws. Over 
tightening will damage the threads of the 
brass inserts.
Install Therapy Intensity Control Knob on the 
front side of Therapy System Top.
Re-assemble Therapy System referring to the 
appropriate sections of this manual for proper 
instructions.

1.
2.

1.

2.
3.
4.

5.

6.

7.

1.

2.

3.

FIGURE 6.25

FIGURE 6.26

CONTRAST KNOB

REMOVE

RETAINING

SCREWS

FIGURE 6.27

REMOVE

RIBBON CABLE

DISCONNECT THE SYSTEM FROM THE POWER 
SOURCE (OUTLET OR REMOVE BATTERY MODULE IF 
INSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE, 
INSTALLATION, REMOVAL, OR REPLACEMENT 
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND 
POSSIBLE DAMAGE TO SYSTEM.

MAKE CERTAIN THE CONTROL BOARD RIBBON 
CABLE IS INSTALLED WITH THE BLACK STRIP AT THE 
NUMBER “1” ON THE CONTROL BOARD. FAILURE 
TO PROPERLY INSTALL RIBBON CABLE COULD 
CAUSE EXTENSIVE DAMAGE TO THE INTERNAL 
COMPONENTS OF THE SYSTEM WHEN TURNED ON.

Содержание SENIOR SOLUTIONS

Страница 1: ...ISO 13485 CERTIFIED Therapy System SERVICEMANUAL Moving Rehabilitation Foward ...

Страница 2: ......

Страница 3: ...ure 4 6 19 F IFC Interferential Premodulated 2p Figure 4 7 19 G Russian Figure 4 8 20 H Microcurrent Figure 4 9 20 I DC Direct Current Figure 4 10 20 J VMS Burst Figure 4 11 21 K VMS FR Figure 4 12 21 5 TROUBLESHOOTING 23 5 1 THERAPY SYSTEM ERROR MESSAGES 23 5 2 THERAPY SYSTEM TESTING 31 5 3 VISUAL INSPECTION 32 5 4 LEAKAGE TESTS 32 5 5 UNIT STARTUP AND FAN TESTING 32 5 6 STIMULATOR TEST SYSTEM SE...

Страница 4: ...PC BOARD ASSEMBLY 76 TOP HOUSING ASSEMBLY 77 SENIOR SOLUTIONS CONTROL BOARD ASSEMBLY 78 CHANNEL 3 4 ELECTROTHERAPY MODULE ASSEMBLY 79 10 SCHEMATICS 80 SENIOR SOLUTIONSTHERAPY SYSTEM CONTROL BOARD 80 82 SENIOR SOLUTIONSTHERAPY SYSTEM ULTRASOUND PC BOARD 83 85 SENIOR SOLUTIONSTHERAPY SYSTEM STIM BOARD 86 95 SENIOR SOLUTIONSTHERAPY SYSTEM CONNECTOR BOARD 96 SENIOR SOLUTIONSTHERAPY SYSTEM CHANNEL 3 4 ...

Страница 5: ......

Страница 6: ...roduced and made available on our web site chattgroup com in lieu of reprinting a complete manual prematurely These service bulletins will provide updated service information and technological improvements to the Senior Solutions Therapy System for use by certified service technicians Due to the complex nature of the technology utilized by Chattanooga Group the recommended troubleshooting techniqu...

Страница 7: ...Stim Board is via a proprietary bus A processor on the Stim Board acts on messages passed to it by the Control Board to set up waveforms and adjust output amplitude Information can likewise be passed from the Stim Board back to the Control Board for monitoring Current Microcurrent Probe Contact Quality indication etc If the Stim Board does not respond as expected to a command from the Control Boar...

Страница 8: ...raphy sEMG Module utilizes a PC board to communicate to the Stim and Control Boards via direct PC Board Contacts The sEMG module reads and transmits muscle activity through lead wires and electrodes The sEMG Module communicates muscle activity data to the Control Board which can store the data on an sEMG Data Card via the onboard Card Reader Writer for viewing on a PC in graph form via the optiona...

Страница 9: ...chnician of possible hazards resulting in the electrical charge disbursement from certain components if handled or serviced improperly 2 1 PRECAUTIONARY DEFINITIONS A Caution B Warning C Danger D Dangerous Voltage E Biohazardous Materials F Corrosive Text with a CORROSIVE indicator will explain possible safety infractions if the chemical components of the battery are exposed to air skin or other m...

Страница 10: ...connect the equipment to an outlet on a different circuit from that to which the other device s are connected and consult the factory field service technician or the Chattanooga Service Department for help Nylatex Wraps contain dry natural rubber and may cause allergic reactions in patients with allergies to latex Ultrasound should be routinely checked before each use to determine that all control...

Страница 11: ...n of electrical current into the heart may cause cardiac arrhythmia Stimulation should not be applied over swollen infected and inflamed areas or skin eruptions e g phlebitis thrombophlebitis varicose veins etc Stimulation should not be applied over or in proximity to cancerous lesions Output current density is related to electrode size Improper application may result in patient injury If any ques...

Страница 12: ...pecific nomenclature of the optional modules TWO CHANNEL COMBINATION THERAPY SYSTEM REFER TO PAGE 8 TWO CHANNEL ELECTROTHERAPY SYSTEM REFER TO PAGE 9 CHANNEL 3 4 ELECTROTHERAPY MODULE REFER TO PAGE 10 OR NIMH BATTERY MODULE REFER TO PAGE 11 CHANNELS 1 2 AND 3 4 OPERATOR REMOTE REFER TO PAGE 16 THERAPY SYSTEM CART REFER TO PAGE 15 DUAL CHANNEL SEMG MODULE REFER TO PAGE 14 INSTALLED TO BOTTOM OF THE...

Страница 13: ...el Serial Decal Two Channel Combo System Ultrasound Applicator 5 cm2 shown User Interface Screen and Buttons Front Access Panel Patient Data Card and sEMG Data Card access port Multimedia Card MMC access port 1 2 3 4 5 6 7 8 9 10 11 12 Front Access Panel Lanyard Operator Remote Control Connector Patient Interrupt Switch Connector Channel 1 Lead Wire Connector Channel 2 Lead Wire Connector Microcur...

Страница 14: ...onnector Patient Interrupt Switch Connector Channel 1 Lead Wire Connector Channel 2 Lead Wire Connector Microcurrent Probe Connector Therapy System to Module Ribbon Cable 12 13 14 15 16 17 18 18 3 1 SENIOR SOLUTIONS THERAPY SYSTEM CONTINUED C Senior Solutions Therapy System Electrotherapy Systems The nomenclature graphics below Figure 3 3 indicate the general locations of the exterior components o...

Страница 15: ...ule to System Mounting Holes 4 Module to System Feet Alignment Indents 5 Power Cord Routing Port 6 Module to System Connector 7 Operator Remote Control Connector 8 Patient Interrupt Switch Connector 9 Channel 3 Lead Wire Connector 10 Channel 4 Lead Wire Connector 11 Microcurrent Probe Connector Also Included Four 4 mm X 20 mm mounting screws Channel 3 and 4 Lead Wires Sample of Dura Stick II Elect...

Страница 16: ...r service to the Senior Solutions Therapy System NiMH Battery Module 3 1 SENIOR SOLUTIONS THERAPY SYSTEM CONTINUED E Senior Solutions Therapy System NiMH Battery Module 1 2 3 4 6 5 1 NiMH Battery Module 2 Extended Front Access Panel 3 Module to System Mounting Holes 4 Module to System Feet Alignment Indents 5 Power Cord Routing Port 6 Module to System Connector NOTE The NiMH Battery Module is not ...

Страница 17: ... service to the Senior Solutions Therapy System Dual Channel sEMG Module 1 sEMG Module Top Housing 2 Module Removal Slot 3 Module to System Mounting Tabs 4 Module to System PC Board Contacts 5 Module to System Retaining Tab 6 sEMG Module Bottom Housing NOTE The Senior Solutions Dual Channel sEMG Module is not operable unless it is connected to the Senior Solutions Therapy System 1 2 3 4 5 6 3 1 SE...

Страница 18: ...nd their functions before performing any operation of or service to the Senior Solutions Therapy System Cart 1 Cart Top 2 System to Cart Retaining Screw 4 3 Storage Bins 6 4 Cart Rear Swivel Casters 5 Cart Base 6 Cart Front Swivel Locking Casters 7 Cart Bottom Access Plate 8 Front and Rear Cart Extrusions 1 2 3 4 5 6 8 7 FIGURE 3 9 3 1 SENIOR SOLUTIONS THERAPY SYSTEM CONTINUED G Senior Solutions T...

Страница 19: ...se Intensity Button 4 Channel 2 Decrease Intensity Button 5 Manual Stimulation Button 6 Channel 1 Decrease Intensity Button 7 Channel 1 Increase Intensity Button NOTE The Senior Solutions Operator Remote Control is not operable unless it is properly connected to the Senior Solutions Therapy System Blue Button for Channels 1 2 Operator Remote Control Orange Button for Channels 3 4 Operator Remote C...

Страница 20: ...ENT INTERRUPT SWITCH CHANNEL 1 LEAD WIRES CHANNEL 2 LEAD WIRES MICROCURRENT PROBE ULTRASOUND APPLICATOR HOME CLINICAL RESOURCES LIBRARY BACK CONTRAST CONTROL 3 2 SENIOR SOLUTIONS THERAPY SYSTEM HARDWARE AND SOFTWARE SYMBOL DEFINITIONS A Hardware Symbols MOVE UP MOVE DOWN MOVE RIGHT MOVE LEFT ACCEPT AND RETURN DO NOT ACCEPT AND RETURN B Software Symbols 1 Operator Remote Control Symbols M INCREASE ...

Страница 21: ...d Electrotherapy Unit 12 750 in 32 4 cm Height Combination and Electrotherapy Unit 8 750 in 22 2 cm Standard Weight Two Channel Combination Unit 7 lbs 3 2 kg Two Channel Electrotherapy Unit 6 lbs 2 7 kg Power Combination and Electrotherapy Units Input 100 240 V 218 75 VA 50 60 Hz Output Internal Power Supply 24 V 7 3 A Duty Cycle Continuous Electrical Class CLASS I Electrical Type Ultrasound TYPE ...

Страница 22: ...ll even at relatively high intensities Output Mode Electrodes Output Intensity 0 110 mA Phase Duration Adjustable 20 1 000 μsec Frequency 1 250 Hz Mode Selection CC or CV Burst Frequency 0 10 bps Frequency Modulation 0 250 Hz Amplitude Modulation Off 40 60 80 and 100 Treatment Time 1 60 min The specifications found in this section provide the necessary waveform specifications to aid in troubleshoo...

Страница 23: ...ationOff 40 60 80 and 100 Treatment Time 1 60 min D High Voltage Pulsed Current HVPC Figure 4 5 The High Voltage Pulsed Current HVPC has a very brief pulse duration characterized by two distinct peaks delivered at high voltage The waveform is monophasic current flows in one direction only The high voltage causes a decreased skin resistance making the current comfortable and easy to tolerate Output...

Страница 24: ...30 10 50 Frequency 1 200 pps Ramp 0 5 sec 1 sec 2 sec and 5 sec Treatment Time 1 60 min Available on Channels 1 2 3 4 Option F IFC Interferential Premodulated 2p Figure 4 7 Premodulated Current is a medium frequency waveform Current comes out of one channel two electrodes The current intensity is modulated it increases and decreases at a regular frequency the Amplitude Modulation Frequency Output ...

Страница 25: ...nt Figure 4 9 Microcurrent is a monophasic waveform of very low intensity that closely simulates the electrical current generated by the human body Microcurrent can be applied via electrodes or probe Output Mode Electrodes or Probe Output Intensity 0 1000 μA Polarity Positive Negative or Alternating Treatment Time 1 60 Min Available on channels 1 2 3 4 Option I DC Direct Current Figure 4 10 DC Cur...

Страница 26: ... 20 10 30 10 50 Frequency 1 200 pps Ramp 0 5 sec 1 sec 2 sec 5 sec TreatmentTime 1 60 min Available on Channels 1 2 3 4 Option 4 2 ELECTROTHERAPY WAVEFORM SPECIFICATIONS CONTINUED VMS FR is a version of the existing VMS waveform that incorporates channel interaction and amplitude modulation to stimulate the movement and coordination of an agonist antagonist muscle group Because the pulse is relati...

Страница 27: ...0 4 Treatment Time 1 30 Mins Head Warming Feature The Head Warming feature of the Senior Solutions Combination Therapy System utilizes ultrasound output resulting in warming of the Sound Head to increase patient comfort With Head Warming enabled ultrasound is emitted without pressing the START button The Applicator LED will not illuminate during the Head Warming period US Channel will indicate Hea...

Страница 28: ...Attempting to delete factory set Sequence 2 Attempting to delete Clinical Protocol Cannot delete factory set Clinical Protocols or factory set Sequences 108 Warning Attempting to save additional User Protocols or Sequences after system memory has reached the maximum allowed 200 Delete some User Protocols or Sequences Refer to appropriate section of the User Manual for instructions 109 110 111 Warn...

Страница 29: ...to use a Non Patient Data Card 3 No Patient Data Card inserted into system port 4 Unknown type of smart card inserted into system A Use a Patient Data Card with proper treatment data for the system B Properly insert a Patient Data Card C Insert a known good Patient Data Card D If problem persists insert a known good Patient Data Card If problem continues contact dealer or Factory for service 123 W...

Страница 30: ...tware to latest version Contact dealer or Chattanooga Group for latest software upgrade and instructions 140 Warning MMC Software upgrade warning Upgrade MMC Software to latest version Contact dealer or Chattanooga Group for latest software upgrade and instructions 141 Warning Battery Module Software upgrade warning Upgrade Battery Software to latest version Contact dealer or Chattanooga Group for...

Страница 31: ... from the system 210 Error Error Calibrating Ultrasound Applicator A Connect a known good Ultrasound Applicator B Replace Ultrasound Board C Replace Control Board 211 Error Error saving Calibration Data to Ultrasound Applicator A Connect a known good Ultrasound Applicator B Replace Ultrasound Board C Replace Control Board 212 Error Ultrasound Applicator not calibrated OK Error A Connect a known go...

Страница 32: ...orming a Software upgrade A TurnTherapy System Off and back On Reattempt upgrade B Replace the PC Board that the software is attempting to upgrade 232 Error Error while performing a Software upgrade A TurnTherapy System Off and back On Reattempt upgrade B Replace the PC Board that the software is attempting to upgrade 233 Error Error while performing a Software upgrade A TurnTherapy System Off and...

Страница 33: ... Board B On Channel 3 4 Electrotherapy Module make certain theTherapy System to Module Ribbon Cable is seated in theTherapy System and on the Module C If problem persists replace appropriate Stim PC Board D Replace Control Board 304 Critical Error Channel A uP on Stim Board reset occurred A OnTherapy System make certain internal Ribbon Cable is seated on Stim PC Board and Control Board B On Channe...

Страница 34: ... Board A Make certain Ultrasound PC Board is completely seated on internal Header and Ribbon Cable is seated on Stim PC Board and Control Board B If problem persists replace Ultrasound PC Board C Replace Control Board 313 Critical Error Error reading from Ultrasound Board A Make certain Ultrasound PC Board is completely seated on internal Header and Ribbon Cable is seated on Stim PC Board and Cont...

Страница 35: ...pletely in the System housing Check 1 4Turn Screw for proper installation B Replace with known good sEMG Module C Replace Stim Board D Replace Control Board Operator Remote Control doesn t work according to User Manual instructions 1 Early Revision of Stim Board and Connector Board 2 Bad Operator Remote Control 3 Bad Connector PC Board 4 Bad Stim Board 5 Bad Control Board A Follow Amendment to Ope...

Страница 36: ... only by authorized technicians trained and certified by Chattanooga Group C Equipment Required Oscilloscope and Probes ESTI 2 Load Test Fixture Digital Multimeter Microcurrent Probe Accessory Patient Interrupt Switch Optional Accessory Operator Remote Control Optional Accessory 1 2 3 1 2 3 4 1 2 3 4 5 6 Senior Solutions Therapy System Ultrasound Applicators Accessories Dielectric Withstand Hi Pot...

Страница 37: ...wing out See Figure 5 2 B Test Results 1 Unit will not START Unit Failed Test a Bad Fuse b Possible bad Main Power Switch c Possible bad Power Supply d Possible bad power outlet or Power Cord 2 Home Screen does not display Unit Failed Test a Possible bad display b Possible bad Control Board c Possible bad Power Supply Visually check power LED LED should illuminate Blue Turn system off with power s...

Страница 38: ...Channels 1 and 2 on the system or Channels 3 and 4 on the Channel 3 4 Electrotherapy Module See Figure 5 3 Connect Lead Wires from the system to the ESTI 2 Load Test Fixture Channel 1 or 3 to Channel 1 IN and Channel 2 or 4 to Channel 2 IN See Figure 5 4 Connect Scope Probes to the Channel 1 To SCOPE and Channel 2 To SCOPE Tabs on the ESTI 2 Load Test Fixture respectively See Figure 5 4 Place ESTI...

Страница 39: ...and press the Up Arrow until 300 is displayed just below Phase Duration Press the Accept and Return Arrow Turn Therapy Intensity Control clockwise until 200 is displayed Press START Compare waveform on scope to Figure 5 5 Press STOP Button Then press Home Button Press Select Channel Button until Channel 2 is selected Repeat steps 2 through 11 B VMS Mode Test Results Waveform is the same between sc...

Страница 40: ...ayed below timer and beside Channel 1 and 2 icons drops to zero 0 Verify that PAUSE is displayed beside the Channel 1 and Channel 2 icons Press STOP button Press the Home Button Move Lead wires to Channels 3 and 4 on 4 Channel Systems and repeat steps 2 through 9 B Interferential Mode Test Results 1 Waveform is the same between scope and Figure 5 6 amplitude dropped to zero when PAUSE and PAUSE di...

Страница 41: ...d 2p button Rotate Intensity Control clockwise until 50 is displayed Press START Compare waveform form on scope to Figure 5 7 Press STOP Then press Home button and move scope probes to Channel 2 and repeat steps 2 through 6 Repeat test on channels 3 and 4 on 4 Channel Systems B Premodulated Mode Test Results Waveform is the same between scope and Figure 5 7 Unit passed test No waveform or consider...

Страница 42: ... and Patient Switch for Ch 1 and 2 was pressed Press any button to continue message appears See Figure 5 9A Press any button Press STOP button Press Home button and move scope probes to Channel 3 and 4 on 4 Channel Systems Repeat steps 2 through 11 B Russian Mode Test Results Waveform is the same between scope and Figure 5 9 amplitude reached 100 and patient switch message displayed when switch pr...

Страница 43: ...inue to ramp and rotate due to Alternating Polarity being selected This is normal Rotate Therapy Intensity Control until 1000 is displayed Press START Compare waveform on scope to Figure 5 11 and Figure 5 11A NOTE The output will alternate between positive and negative on the scope Press STOP button and then press Home button Select the next channel to be tested by pressing the Select Channel butt...

Страница 44: ... displayed in Polarity icon Compare waveform form on scope to Figure 5 13 Press Display button until Peak Current is displayed in the Display icon The numbers displayed for amplitude must not exceed 1 5 Amps See Figure 5 14 Press STOP B High Voltage Pulsed Current HVPC Mode Test Results Waveforms on scope the same as Figures 5 12 and 5 13 and Amps do not exceed 1 5 Unit passed test No waveform or ...

Страница 45: ...be heard this is the Search Mode beep A Contact Quality Scale will also appear on the screen Touch the probe to the metal tip of the Black Leadwire See Figure 5 16 The beep search mode should increase in speed and the Contact Quality scale should display a full vertical green bar on Color Systems gray on Monochrome See Figure 5 17 Press the blue button on the probe The Search Mode beep should stop...

Страница 46: ...ate Stim Board Good Search Mode beep but no beep speed increase when probe touched to metal tip of the Black Leadwire Unit failed test Replace probe and lead wire with known good probe and lead wire Repeat test and use process of elimination to determine if probe or lead wire is the problem Same results after test is repeated replace appropriate Stim Board Tones OK but no green or gray vertical ba...

Страница 47: ...cator LED while connecting to System The LED should flash green five times Look at the Ultrasound channel icon It should read Available See Figure 5 19 Press the Ultrasound button Press the Edit Button Press the Head Warming button until On is displayed Press the Back button Turn System Off and back On with Main Power Switch After System boots view the Ultrasound icon Head Warming should be visibl...

Страница 48: ... MHz is displayed within the Frequency icon Repeat test and compare readings to Figure 5 22 NOTE The Applicator LED should constantly illuminate green during the Applicator Output tests B Ultrasound Applicator Output Test Results Output ranges fall within the specified ranges as listed in Figure 5 22 Unit passed test Readings fall outside specified ranges of Figure 5 22 a Possible bad degassed wat...

Страница 49: ...s shown in Figure 5 24 Press Frequency button until 3 3 MHz is displayed within the Frequency icon Repeat test and compare readings to Figure 5 24 B Ultrasound Duty Cycle Test Results Duty Cycles fall within the specified ranges as listed in Figure 5 24 Unit passed test Readings fall outside specified ranges of Figure 5 24 a Possible bad degassed water in Power Meter b Possible use of Power Meter ...

Страница 50: ...lect Waveform button Press the Up or Down Arrow button until IFC 4p is highlighted Press the Accept and Return button Press Edit Stim button and rotate Intensity Control clockwise until 50 50 mA is displayed in Amplitude Ch 1 2 icon Press START Touch the Ultrasound Applicator to the Combo Contact on the ESTI 2 Load Test Fixture The Combo Indicator on the ESTI 2 should illuminate See Figure 5 26 Co...

Страница 51: ...EMG Lead Wires B sEMG Test Procedures Refer to Figure 5 29 to set up the Audio Signal Generator a Plug the Audio Signal Generator Test Leads into Generator SYNC Ports b Set the FREQ RANGE Hz to X1 c Turn the amplitude knob up to maximum d Set the WAVEFORM to Sine waveform e Set the ATTEN to O f Set the FREQUENCY DIAL to 100 g Turn Audio Signal Generator On Turn System On View Home screen for the p...

Страница 52: ...into the Attenuator Make certain each sEMG Lead is connected to its respective color on the Attenuator See Figure 5 32 Press the sEMG button on Home screen Channel 1 should read 7 or less See Figure 5 33 If Channel 1 reads less than 7 repeat steps 2 through 5 on Channel 2 If any Channel being tested reads greater than 7 replace the sEMG Module and re test NOTE The reading on the Channel not being ...

Страница 53: ...sted reads below 604 or greater than 738 replace the respective sEMG Module and re test NOTE The reading on the channel not being tested may vary This is insignificant as it is not under load C sEMG Test Results If any sEMG Channel fails any part of the tests as described in 5 19 B steps 2 through 9 then the module fails the test Make certain the sEMG Module is completely seated in system housing ...

Страница 54: ...Accept and Return Arrow button Press the Edit Stim button Rotate Treatment Intensity Knob until 5 0 mA CC is displayed Then press the Back Button Press START sEMG Stim Button The Audio Signal Generator and the attenuator should trigger the stim function of the Therapy System and Running will display in the selected channel See Figure 5 35A E sEMG Stim Test Results Stim function is triggered System...

Страница 55: ...Charge Level icon will be fully black If battery is fully charged perform steps 4 through 7 below then perform step 3 Allow battery to charge until it can be verified that it is charging by viewing the Charge Level icon until the level indicator fills more of the Charge Level icon Set up the Ultrasound Test as described in 5 16 Set Duty Cycle to 100 Set Display to Watts rotate Therapy Intensity kn...

Страница 56: ...iMH Battery Module Checks Results Should any check fail replace the entire module No component parts are or will be made available for the Battery Module by Chattanooga Group If the problem persists replace the Control Board 10 11 12 13 1 2 FIGURE 5 40 FIGURE 5 39 FIGURE 5 38 NiMH batteries contain Class E corrosive materials In the event of battery cell rupture or leakage handle battery module we...

Страница 57: ...Cable and unroll Ribbon Cable See Figure 6 4 NOTE When turning over Therapy System onto its face place a clean soft cloth under the lens to prevent scratching or lens damage If the system is equipped with an sEMG Module leave it in place to maintain sEMG functions for Channels 1 and 2 The sEMG Module will not interfere with installation of a Module to the Therapy System 1 2 3 1 2 3 4 5 6 FIGURE 6 ...

Страница 58: ...in Figure 6 6 Align Therapy System Feet with Module Indentions RIBBON CABLE MUST BE POSITIONED AS SHOWN Place System onto Module making certain the System Feet are within the Module Identions 7 8 9 DISCONNECTTHE SYSTEM FROMTHE POWER SOURCE OUTLET OR REMOVE BATTERY MODULE IF INSTALLED BEFORE ATTEMPTING ANY MAINTENANCE INSTALLATION REMOVAL OR REPLACEMENT PROCEDURESTO PREVENT ELECTRICAL SHOCK AND POS...

Страница 59: ...dule to the System with four 4 mm x 20 mm Screws See Figure 6 7 Tighten only enough to prevent Module from shifting on Therapy System Set the assembly upright on the work surface and install the new Extended Front Access Panel to the Lanyard See Figure 6 8 NOTE When mounting the Front Access Panel to the Therapy System make certain the Lanyard does not become kinked Route Power Cord through Module...

Страница 60: ... Module in reverse order of installation beginning with step A 13 After module is removed follow instructions in steps A 14 and A 15 13 14 15 1 2 Verify that the module installed is the module displayed in the message BEFORE pressing the START Button If it is not DO NOT press the START Button Turn the system OFF and back ON If the problem persists call the dealer or Chattanooga Group Technical Sup...

Страница 61: ...rotherapy or Combination Therapy System An sEMG Module cannot be installed on the NiMH Battery Only two channels of sEMG can be used at any given time even if the Therapy System is equipped with two sEMG Modules Disconnect Power Cord from the power source and turn system over on its face NOTE When turning over Therapy System onto its face place a clean soft cloth under the lens to prevent scratchi...

Страница 62: ...ition See Figure 6 14 Re install module if required Refer to 6 2 part C Route Power Cord through the Module and connect to the Therapy System Install all cables Lead Wires etc to the Therapy System and Module where applicable Refer to page 15 for symbol definitions See Figure 6 15 3 4 5 6 DISCONNECTTHE SYSTEM FROMTHE POWER SOURCE OUTLET OR REMOVE BATTERY MODULE IF INSTALLED BEFORE ATTEMPTING ANY M...

Страница 63: ...28027 to protect and cover the Stim Board contacts The Therapy System may be placed back into service without an sEMG Module installed as long as the sEMG Plug Kit is installed to protect the PC Board contacts Refer to steps 7 and 8 to finalize removal 7 8 1 2 3 FIGURE 6 16 FIGURE 6 17 Verify that the module installed is the module displayed in the message BEFORE pressing the START Button If it is...

Страница 64: ...bottom See Figure 6 18 Turn System over on its feet and carefully separate the System Top from the Bottom Housing NOTE On Combination Systems it may be necessary to use a Flat Blade Screwdriver around the Ultrasound Applicator Rest to separate the top from the bottom See Figure 6 19 Raise the System Top and disconnect the Control Board Ribbon Cable from the Stim Board or Ultrasound Board on Combin...

Страница 65: ...lacing Top to Bottom Replace System Top by reversing the steps 1 5 on page 59 NOTE Do not overtighten the screws Overtightening will damage the threads of the brass inserts 6 DISCONNECTTHE SYSTEM FROMTHE POWER SOURCE OUTLET OR REMOVE BATTERY MODULE IF INSTALLED BEFORE ATTEMPTING ANY MAINTENANCE INSTALLATION REMOVAL OR REPLACEMENT PROCEDURESTO PREVENT ELECTRICAL SHOCK AND POSSIBLE DAMAGETO SYSTEM ...

Страница 66: ...ee Figure 6 23 Remove the Fan Baffle from the Fan Housing See Figure 6 24 C Replacing Fan Replace new Fan part number 27158 by reversing the steps 1 3 above NOTE Do not over tighten the screws Over tightening will damage the threads of the brass standoffs 1 2 3 4 1 FIGURE 6 22 FIGURE 6 23 REMOVE SCREWS REMOVE FAN HARNESS FIGURE 6 24 REMOVE BAFFLE 61 DISCONNECTTHE SYSTEM FROMTHE POWER SOURCE OUTLET...

Страница 67: ...om the Control Board Assembly C Replacing Control Board Assembly Replace new Control Board Assembly part number 27053 by reversing the steps 1 7 above NOTE Do not over tighten the screws Over tightening will damage the threads of the brass inserts Install Therapy Intensity Control Knob on the front side of Therapy System Top Re assemble Therapy System referring to the appropriate sections of this ...

Страница 68: ...g Keymat Assembly Replace with new Keymat Assembly Kit by reversing the steps 1 6 above NOTE Do not over tighten the screws Over tightening will damage the threads of the brass inserts Re assemble Therapy System referring to the appropriate sections of this manual for proper instructions 1 2 1 2 3 4 5 6 1 2 FIGURE 6 28 FIGURE 6 29 REMOVE RETAINING SCREWS LIFT PC BOARD AND RELEASE PLASTIC CLIPS FIG...

Страница 69: ...ctor Board are facing up Make certain Connector Board is completely seated in Stim Board Connector See Figure 6 33 Re install the Connector Infill and Lanyard Refer to Figure 6 31 Re assemble Therapy System Refer to 6 4 part C for proper instructions NOTE Do not over tighten the screws Over tightening will damage the threads of the brass inserts 1 2 3 1 2 3 FIGURE 6 31 FIGURE 6 32 REMOVE CONNECTOR...

Страница 70: ...is clear of the Stand Off Barbs See Figure 6 35 Use the Needle Nose Pliers to release the Ultrasound Board from the front Stand Offs See inset in Figure 6 34 C Replacing Ultrasound Board Replace new Ultrasound Board part number 27055 in reverse order of steps 1 5 above Re assemble Therapy System Refer to 6 8 and 6 4 part C for proper instructions NOTE Do not over tighten the screws Over tightening...

Страница 71: ...C Replacing Stim Board Replace new Stim Board part number 27056 in reverse order of steps 1 8 above Make certain Jumpers are properly set on Stim Board See Figure 6 38A NOTE Position Ribbon Cable so the Blue Strip is positioned as shown in Figure 6 37A Press on Stim Board while seating Ribbon Cable to the system On Combination Systems install the Ribbon Cable and turn fastener before installing th...

Страница 72: ... securing the Power Supply Assembly to the System Housing See Figure 6 39 Lift Power Supply Assembly up to remove from rear mounting tabs See Figure 6 40 Using the Digital Multimeter discharge the Power Supplies as follows a 100 Watt Power Supply Discharge Capacitor C4 See Figure 6 41 b 75 Watt Power Supply Combo Systems Only Discharge Capacitor on back of Power Supply PC Board Pins are on either ...

Страница 73: ...m the inside top of the Power Supply See Figure 6 43 NOTE Steps 10 and 11 below apply only to Combination Systems Remove 75 Watt Power Supply from Mounting Bracket by removing the two mounting screws on the back of Mounting Bracket securing the 75 Watt Power Supply See inset at Figure 6 43 Using Insulated Needle Nose Pliers remove the 100 Watt Power Supply Harness from the 75 Watt Power Supply See...

Страница 74: ...er of steps 1 4 above Make Certain the words THIS SIDE UP on Connector Board are facing up Make certain Connector Board is completely seated in Stim Board Connector See Figure 6 47 Re install the Connector Infill and Lanyard Refer to Figure 6 46 Re assemble Therapy System referring to the appropriate sections of this manual for proper instructions NOTE Do not over tighten the screws Over tightenin...

Страница 75: ...6 48A C Replacing Channel 3 4 Electrotherapy Module Stim Board Replace new Stim Board part number 27056 in reverse order of steps 1 4 above Re assemble Module and Therapy System referring to the appropriate sections of this manual for proper instructions NOTE Do not over tighten the screws Over tightening will damage the threads of the brass inserts 1 2 1 2 3 4 1 2 FIGURE 6 48 CONNECTOR BOARD FIGU...

Страница 76: ...em to the cart by tightening the screws by hand only Do not use a wrench to tighten the screws Over tightening may cause damage to the System or Module housing Plug Power Cord into the System Mains Disconnect and reinstall the Rear Access Panel Install all lead wires and cables to the System Install Storage Bins into Therapy System Cart START with bottom Storage Bin first B Dismounting Therapy Sys...

Страница 77: ...sh com on the internet Follow the instructions as given by NOVUS on their product Do not use alcohol or chlorine based solvents as this may damage the lens 7 2 CALIBRATION REQUIREMENTS Ultrasound Applicators Annual factory calibration is required for all Ultrasound Applicators Only the Applicators should be sent to the factory for this procedure 7 3 FIELD SERVICE A All field service procedures as ...

Страница 78: ...r Meter Press the Ultrasound Calibration button See Figure 8 2 Press the Head Size button until the size applicator being calibrated is displayed See Figure 8 3 Press the START button refer to Figure 8 3 Follow the instructions displayed on the Therapy System Repeat this procedure for each Ultrasound Applicator associated with the Therapy System being serviced 1 2 3 1 2 3 4 5 6 FIGURE 8 1 FIGURE 8...

Страница 79: ...OP ASSEMBLY BOTTOM ASSEMBLY 7 ITEM NUMBER PART NUMBER DESCRIPTION QTY REQ D 1 27159 Bottom Assembly to Top Assembly Ribbon Cable 1 2 27306 Rear Access Panel 1 3 27138 Screw M3 x 16 mm 4 4 27029 Front Access Panel 1 5 27020 Lanyard 1 6 21188 Screw 4 40 x 375in 1 7 27007 Front Infill 1 ...

Страница 80: ...TY REQ D 1 27022 Mains Harness Clip 1 2 27276 Main Power Switch 1 5 27277 Mains Power Input Connector 1 6 27010 Applicator Holder 1 7 27274 Feet 4 8 27436 Stim Board Turn Pin 1 9 27059 Stim Connector Board 1 10 27142 Screws M3 x 6mm 2 11 27006 Base Housing 1 12 27048 100 Watt Power Supply 1 13 27049 75 Watt Power Supply Combination Systems Only 1 14 27000 Power Supply Mounting Bracket 1 ...

Страница 81: ...em 9 PARTS 1 2 3 3 4 3 COMBINATION STIM ULTRASOUND PC BOARD ASSEMBLY ITEM NUMBER PART NUMBER DESCRIPTION QTY REQ D 1 27055 Ultrasound PC Board 1 2 27161 Ultrasound to Stim Header 1 3 27160 PC Board Stand Off 4 4 27057 Stim PC Board 1 ...

Страница 82: ...BLY SEE PAGE 63 79 81 5 5 6 5 7 ITEM NUMBER PART NUMBER DESCRIPTION QTY REQ D 1A 28122 Top Housing Blue 1 1B 28400 Lens Senior Solutions 1 2 27026 Left Keymat 1 3 27027 Right Keymat 1 4 27051 Keymat PC Board 1 5 27142 Screw M3 x 6 mm 10 6 27028 Bottom Keymat 1 7 27305 Intensity Control Knob 1 1B ...

Страница 83: ...splay Mounting Bracket Spacer 2 2 27162 Color Display 1 3 27012 Contrast Knob Not Functional 1 4 27053 Color Control PC Board 1 5 27137 Screw M4 x 8 mm 2 6 21188 Screw PSM PT Type 4 7 27021 Air Exhaust Funnel 1 8 27158 Cooling Fan 1 9 27136 Screw 4 40 x 375 in 4 10 27144 Fan Mounting Stand Off M4 x 16 mm 2 11 27024 MMC Card Top 1 12 27009 Patient Data Card and MMC Card System Port Housing 1 ...

Страница 84: ... PART NUMBER DESCRIPTION QTY REQ D 1 27015 Module Top 1 2 27353 Stim Board to Ribbon Cable Header 1 3 27057 Stim PC Board 1 4 27016 Module Bottom Housing 1 5 27150 Feet 4 6 27061 Channel 3 4 Stim Connector PC Board 1 7 27017 Connector Infill 1 8 27136 Module to Therapy System Ribbon Cable Mounted onto Therapy System 1 ...

Страница 85: ...80 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONTROL BOARD 27053 Sheet 1 ...

Страница 86: ...81 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONTROL BOARD 27053 Sheet 2 ...

Страница 87: ...82 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONTROL BOARD 27052 Sheet 3 ...

Страница 88: ...83 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM ULTRASOUND PC BOARD 27055 Sheet 1 ...

Страница 89: ...84 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM ULTRASOUND PC BOARD 27055 Sheet 2 ...

Страница 90: ...85 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM ULTRASOUND PC BOARD 27055 Sheet 3 ...

Страница 91: ...86 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 1 ...

Страница 92: ...87 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 2 ...

Страница 93: ...88 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 3 ...

Страница 94: ...89 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 4 ...

Страница 95: ...90 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 5 ...

Страница 96: ...91 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 6 ...

Страница 97: ...92 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 7 ...

Страница 98: ...93 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 8 ...

Страница 99: ...94 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 9 ...

Страница 100: ...95 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 10 ...

Страница 101: ...96 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONNECTOR BOARD 27059 ...

Страница 102: ...97 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CHANNEL 3 4 ELECTROTHERAPY MODULE CONNECTOR BOARD 27061 ...

Страница 103: ...98 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM POWER SUPPLIES 27048 100 Watt Power Supply 27049 75 Watt Power Supply Combination Systems Only ...

Страница 104: ...99 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM LASER MODULE BOARD 27069 ...

Страница 105: ......

Страница 106: ......

Страница 107: ...selling dealer or a certified Company service technician Any malfunction or failure in the Product caused by product misuse including but not limited to the failure to provide reasonable and required maintenance or any use that is inconsistent with the Product Users Manual COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES Some locations do not allow the exclusion or ...