19
Senior Solutions™ Therapy System
4 SPECIFICATIONS
E. VMS™- Figure 4.6
VMS is a symmetrical biphasic waveform with a 100 μsec interphase
interval. Because the pulse is relatively short, the waveform has a low
skin load, making it suitable for applications requiring high intensities,
such as in muscle strengthening protocols.
Output Mode………………………………………Electrodes
Output Intensity………………………………………0-200 mA
Channel Mode……………………Single, Reciprocal, Co-Contract
Phase Duration………………………………………20-400 μsec
Mode Selection…………………………………………CC or CV*
Anti-Fatigue……………………………………………Off or On
Set Intensity…Individual Channel Intensity Setting in Reciprocal and
…………………………………………………Co-Contract modes
Cycle Time…………Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, 10/50
Frequency…………………………………………………1-200 pps
Ramp………………………………0.5 sec, 1 sec, 2 sec, and 5 sec
Treatment Time…………………………………………1-60 min
Available on Channels.……………………… 1 & 2, 3 & 4 Option
F. IFC (Interferential) Premodulated (2p)- Figure
4.7
Premodulated Current is a medium frequency waveform.
Current comes out of one channel (two electrodes). The
current intensity is modulated: it increases and decreases
at a regular frequency (the Amplitude Modulation
Frequency).
Output Mode……………………………………Electrodes
Output Intensity……………………………………0-100 mA
Carrier Frequency…………………………………2,500 Hz
Beat Fixed (Sweep Off )……………………………1-200 Hz
Sweep Low Beat Frequency………………………1-199 Hz
Sweep High Beat Frequency…………………… 2-200 Hz
Cycle Time…Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, 10/50
Mode Selection ……………………………………CC or CV*
Treatment Time………………………………… 1-60 Min
Available on Channel………………… 1 & 2, 3 & 4 Option
FIGURE 4.6
*CC= Constant Current
CV= Constant Voltage
4.2 ELECTROTHERAPY WAVEFORM SPECIFICATIONS (CONTINUED)
FIGURE 4.7
Содержание SENIOR SOLUTIONS
Страница 1: ...ISO 13485 CERTIFIED Therapy System SERVICEMANUAL Moving Rehabilitation Foward ...
Страница 2: ......
Страница 5: ......
Страница 85: ...80 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONTROL BOARD 27053 Sheet 1 ...
Страница 86: ...81 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONTROL BOARD 27053 Sheet 2 ...
Страница 87: ...82 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONTROL BOARD 27052 Sheet 3 ...
Страница 91: ...86 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 1 ...
Страница 92: ...87 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 2 ...
Страница 93: ...88 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 3 ...
Страница 94: ...89 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 4 ...
Страница 95: ...90 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 5 ...
Страница 96: ...91 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 6 ...
Страница 97: ...92 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 7 ...
Страница 98: ...93 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 8 ...
Страница 99: ...94 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 9 ...
Страница 100: ...95 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM STIM BOARD 27057 Sheet 10 ...
Страница 101: ...96 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM CONNECTOR BOARD 27059 ...
Страница 104: ...99 Senior Solutions Therapy System 10 SCHEMATICS SENIOR SOLUTIONS THERAPY SYSTEM LASER MODULE BOARD 27069 ...
Страница 105: ......
Страница 106: ......