3D ActiveTrac
®
36
TrOuBLEShOOTiNg
• Remove the blown fuse from
the fuse holder and replace it
with a new fuse.
NOTE:
See Specifications
Section for fuse rating. If both
fuses are blown, replace them.
• Gently push fuse holder back
into place on the front of the
Power Control Box ensuring
that the voltage mark is on
top. The holder will snap into
place.
• Gently lift up and snap the
fuse holder cover in place.
Important:
If the unit
continues to blow fuses contact
customer service.
If the traction unit doesn’t work:
• Make sure the unit is getting power.
• Make sure the Patient Control Switch is not physically
damaged or something is not pressed against the button.
• If the pump is running but there is no traction force, check
for air leaks between the pump and cylinders (determined
by hissing sound during operation).
• If the Touchscreen Control Panel will not accept or hold a
command, STOP and call service.
Do not
attempt to use
the unit until the problem is resolved.