3D ActiveTrac
®
29
OpErATiON
TOuChSCrEEN CONTrOL pANEL (TCp) (CONTiNuEd)
To set the treatment time, press the box under “Treatment
Time”. A numeric display will appear. Enter a 2 digit number
by pressing the keys (the number you entered appears in the
upper right corner of the screen). If you are satisfied with the
entry, press the enter arrow (bottom row). If you need to make
a change, press the backspace arrow (top row), then re-enter
the desired number.
The acceptable range for treatment time entries is 1 to 60
minutes.
NUMERIC KEYPAD
Force Set
The Force Set option allows you to enter the maximum
treatment force. In Static mode, the maximum treatment force
is the force for the entire treatment time (excluding ramp up
and ramp down times). In Intermittent mode, the maximum
treatment force is the force during the Hold phase of the Hold/
Rest cycle.
The acceptable range for treatment force entries depends
upon whether you are in cervical or lumbar mode. The range
for cervical mode is 10 to 50 lbs (4 to 23 kg). The range for
lumbar mode is 10 to 200 lbs (4 to 90 kg). The default is 10 lbs
(4 kg).
Start
Pressing Start will begin the treatment and advance to the
Treatment Display screen.
Back
NOTE:
Once Start is pressed you can only press Pause and
Back to set the Force/Time Screen. Pressing Back returns to the
previous screen. Press Back if you wish to change a selection on
a previous screen.