TrOuBLEShOOTiNg
If the Hi/Lo or Flexion/Extension motors don’t work,
make sure the motor is receiving power:
• Check the plug to make sure it is securely plugged into a
specified electrical wall outlet.
• Check the plug to make sure it is securely plugged into the
receptacle at the front of the Power Control Box.
• If the table is plugged in, check building circuit breaker to
make sure wall outlet is receiving electricity.
• Thermal cutout may have activated. Allow the motor to cool.
• The 5 amp control box fuse may have blown.
ONLY
QUALIFIED SERVICE PERSONNEL
should replace as follows:
• Shut off power to the Power Control Box.
• Remove the Power Cord from the wall outlet.
• Remove the Power Cord from the Power Control Box.
• Using a small flat screwdriver,
lightly pry open the fuse
holder cover. The cover will
drop down.
• Then, using the screwdriver,
lightly pry open the fuse
holder.
• Gently pull out the fuse
holder.