4.7.3
Spirit™ CB09 Control Box – Overview
CB09 Control Box
CB09 is common for the Spirit™ and Spirit Plus™
CB09 is unique for the Spirit Select™
Typical connections for all bed
DJB Junction Box
Pendant/Siderail controls and Footboard Staff Control
cable connections are labelled
Not present on Spirit Select™
Communication Cable (8 pin, DIN style, double keyed connector)
Spirit™ and Spirit Plus™ - center cable from DJB Junction Box
Spirit Select™ - cable to COMM header on CB10
Actuator Cable Plug (4 pin, DIN style, double keyed connector)
Typical all beds – COLOUR CODED
Yellow: Hi-Lo Actuator
(Head Mounted)
White: Hi-Lo Actuator
(Foot Mounted)
Red: Head Actuator
Blue: Knee Actuator
DC Power Cable (Ø5.5 mm X Ø2.5 mm male plug end)
Spirit™ and Spirit Plus™ - cable from back-up battery
Spirit Select™ - power cable to DC power input header on CB10
Power Supply Cable (IEC 320 C13 female plug end)
Typical on all beds
Содержание Spirit
Страница 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Страница 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Страница 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Страница 33: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 33 of 110 Section 3 Bed Operation ...
Страница 81: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 81 of 110 Section 4 Care Maintenance ...
Страница 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...