Spirit™ Bed Owner’s Manual – 1-70-001 – Rev D
© Carroll Hospital Group Inc.
Page 74 of 110
3.12
Calibration Menu Navigation
NOTE:
A scale or bed angle calibration attempt may be abandoned at any time. Simply remove the
footboard from the bed and then replace it. The home screen will be displayed when the menu
display is powered up. The scale and bed angle calibration will, respectively, revert to their
previous, successfully completed calibration settings to enable scale functionality.
NOTE:
The scale must be calibrated if any/all of the load cells have been replaced. However, the scale
calibration procedure must be initiated by staff. Scale calibration will automatically be initiated if
the CB10 is replaced.
Turn to
page 80 for
instructions
Turn to
page 76 for
instructions
Turn to
page 75 for
instructions
From page 63
Access Code
Successfully Entered