3.5.3
Bed Exit Feature
The
enhanced
footboard staff control on the
Spirit Select™ is equipped with an
integral bed exit feature to help monitor and report patient activity with audible
and/or Nurse/Priority Call alarms. This feature is not intended to replace patient
monitoring by staff.
The Bed Exit Alarm can be set to three different sensitivity levels corresponding to
the level of monitoring a patient requires.
The Bed Exit Alarm can be set to produce an audible bed alarm (three different
volume settings) or to send a Nurse/Priority Call alarm or it can be set to produce
both audible and Nurse/Priority Call alarms.
Bed Exit Activation, Deactivation, and Sensitivity Level Selection
To activate the Bed Exit Alarm, press the SET/RESET button to cycle through the
alarm sensitivity levels. When the desired alarm sensitivity level has been selected,
the SET/RESET button and the selected sensitivity level icon will remain illuminated
to indicate that the bed exit alarm has been activated. An audible “chirp” will also
sound to indicate that the Bed Exit Alarm has been armed.
To de-activate the Bed Exit Alarm, press the SET/RESET button to cycle through the
three alarm sensitivity levels. When the SET/RESET button and no alarm sensitivity
icons are illuminated the Bed Exit Alarm has been de-activated.
IMPORTANT: The Bed Exit Alarm can ONLY be activated if there is weight on the
bed (20lbs minimum). The Bed Exit Alarm can ONLY be de-activated using the
SET/RESET button.
The factory default alarm chime setting is audible chime (high volume) and Nurse
Call alarm. To change this setting or chime tone see below.
Bed Exit Alarm Volume Selection
To set the desired Bed Exit Alarm volume, press the CHIME button to begin cycling
through the alarm volume settings. There are three alarm volume settings (low,
medium, high) and a Nurse Call setting. Cycling through the alarm volume settings
one time will allow staff to select the desired alarm volume setting OR the
Nurse/Priority Call alarm. Cycling through the alarm volume settings a second time
will allow staff to select the desired alarm volume setting AND the Nurse/Priority
Call alarm.
When the desired chime is selected, the CHIME button and the selected alarm
volume and/or Nurse/Priority Call icon(s) will remain illuminated.
Chime Tone Selection
To change the default chime tone, the Bed Exit Alarm must first be activated. Press
the SET/RESET button to activate the Bed Exit Alarm.
Press and hold the CHIME button for 5 seconds to begin cycling through the
available chime tones. Each chime tone will sound (at the selected volume) for
several seconds and will be followed by a brief pause before the next available
chime tone begins. When the desired chime tone is heard, release the CHIME
button to set. The selected chime tone will sound as confirmation that the chime
tone selection has been changed.
Locking Bed Exit Alarm Settings
To lock the setting of the Bed Exit Alarm, press and hold both the SET/RESET &
CHIME buttons simultaneously for 5 seconds. An audible “chirp” will indicate that
the Bed Exit Alarm setting has been locked.
To unlock the Bed Exit Alarm settings, simply repeat this process.
Содержание Spirit
Страница 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Страница 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Страница 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Страница 33: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 33 of 110 Section 3 Bed Operation ...
Страница 81: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 81 of 110 Section 4 Care Maintenance ...
Страница 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...