Spirit™ Bed Owner’s Manual – 1-70-001 – Rev D
© Carroll Hospital Group Inc.
Page 17 of 110
WARNING
In the event a person is exposed to sulphuric acid, flush contacted area with large
amounts of water for at least 15 minutes. Remove contaminated clothing and obtain
medical attention if necessary. Eye wash and emergency shower should be readily
available. If swallowed, give large amounts of water. DO NOT induce vomiting, obtain
medical treatment.
WARNING
If a sulphuric acid spill is found under the bed, dilute the spill cautiously with five to six
volumes of water and gradually neutralize with sodium bicarbonate, soda ash or lime.
When exposure level is not known, wear NIOSH approved positive pressure self-
contained breathing apparatus. Reference North American Emergency Response
Guidebook #154. Be sure to wear acid resistant gloves, safety glasses, and acid resistant
clothing when cleaning a sulphuric acid spill.
1.9
Grounding Instructions
This electric bed must be grounded.
In the event of a malfunction or breakdown, grounding provides a path of least resistance for electric
current, thereby reducing the risk of electric shock.
This product is equipped with a cord having an equipment-grounding conductor and a grounded plug. The
plug must be inserted into an appropriate electric outlet that is properly installed and grounded in
accordance with all local codes and ordinances.
DANGER
Improper connection of the equipment-grounding conductor can result in electrocution.
Check with a qualified electrician or service person if you are doubtful that the electrical
outlet is properly grounded.
DO NOT modify the three prong plug provided. If it will not fit into the electrical outlet,
have a proper outlet installed by a qualified electrician.
DANGER
Electric components, such as actuators, exposed to water, may cause severe injury
including electrocution.
During routine bed operation or when cleaning, DO NOT expose electrical components to
water. Unplug and de-energize electrical components when water products are near
these components.
CAUTION
Grounding reliability can only be achieved when bed is connected to an equivalent
receptacle marked “HOSPITAL ONLY” or “HOSPITAL GRADE”.
Содержание Spirit
Страница 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Страница 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Страница 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Страница 33: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 33 of 110 Section 3 Bed Operation ...
Страница 81: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 81 of 110 Section 4 Care Maintenance ...
Страница 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...