3.5.2
Patient Weigh Scale
The Spirit Select™ is equipped with an integral scale to facilitate the weighing of patients while in the bed. The
scale is always activated and will display the patient’s weight in either kilograms (kg) or pounds (lbs)
depending upon which unit of measure has been set for patient weight display. The factory default displayed
patient weight is pounds (lbs).
The Spirit Select™ scale is very sensitive. Patient weight will be displayed most accurately if nothing is
touching the bed and the bed is not contacting the wall, patient station, etc.
ACCURACY: The Spirit Select™ scale has an approximate error of 1% of patient weight. This equates to an
approximate error of ±0.8 kg (±1.8 lbs) for a patient weighting 80 kg (176 lbs).
Zeroing the Scale (Establish Tare Value)
1.
Press and hold the ZERO button for 3
seconds
2.
Release the ZERO button when instructed
3.
DO NOT TOUCH BED
4.
Ensure that the displayed weight is 0.0 (lbs
or kg – depending upon which unit has be
set)
IMPORTANT: DO NOT zero the scale with a
patient in the bed. This will result in a zero patient
weight being displayed.
IMPORTANT: Make sure that all necessary
equipment and linens are on the bed prior to
zeroing the scale.
IMPORTANT: If the bed detects any movement
while attempting to zero the scale, a warning
message will be displayed and the scale zeroing
procedure will have to be repeated.
Scale Related Activities
Changing patient weight display units:
•
If setting is not locked - see page 56 for Main Menu navigation and page 58 for instructions
•
If setting is locked - see page 63 for Service Menu navigation and page 65 for instructions
Patient weight change zero: see page 56 for Main Menu navigation and page 57 for instructions
Change equipment on the bed: see page 56 for Main Menu navigation and page 59 for instructions
Adjusting patient display weight: see page 63 for Service Menu navigation and page 64 for instructions
Weight change alarm: see page 63 for Service Menu navigation and page 67 for instructions
NOTE:
Carroll Hospital Group recommends that the bed is zeroed prior to use by each new patient and
patient weight change is zeroed once the new patient enter the bed.
Содержание Spirit
Страница 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Страница 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Страница 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Страница 33: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 33 of 110 Section 3 Bed Operation ...
Страница 81: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 81 of 110 Section 4 Care Maintenance ...
Страница 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...