Spirit™ Bed Owner’s Manual – 1-70-001 – Rev D
© Carroll Hospital Group Inc.
Page 41 of 110
3.4.4
Patient Lock-Outs
Patient lock-outs restrict the patient from initiating head, knee, and bed
elevation from the side rail controls or the pendant. The degree of
restriction depends on the lock-out option selected.
Staff may choose to restrict patient access to one, two, or all three bed
functions. To restrict a function, press the patient lock-out button under
that particular function. The yellow LED in the top right corner of the
button (for Spirit™ and Spirit Plus™) or the or the entire lock-out button
(for Spirit Select™) will illuminate when patient control of a particular
function has been locked-out.
To unlock, simply press the patient lock-out button again. When the patient
lock-out button is no longer illuminated, patient control of that particular
function has been restored.
IMPORTANT: Patient lock-outs also restrict bed operation from both of the
staff rail controls.
NOTE:
When patient lock-outs have been activated, the footboard staff control remains fully active.
3.4.5
Master Lock-Out
To completely restrict access to all bed functions from all control locations
(staff and patient controls), use the Master Lock-out function.
To activate the Master Lock-Out, press and hold all three patient lock-out
buttons at the same time. All three yellow LED’s in the top right corner of
the three patient lock-out buttons (for Spirit™ and Spirit Plus™) or all three
entire patient lock-out buttons (for Spirit Select™) will flash sequentially to
indicate that Master Lock-Out has been activated.
To deactivate the Master Lock-Out function, simply repeat this process.
Press and hold all three patient lock-out buttons at the same time. When all
three patient lock-out buttons are no longer illuminated, Master Lock-Out
has been deactivated and both staff and patient control from all control
locations has been restored.
NOTE:
When Master Lock-Out has been activated, the emergency CPR function and Nurse Call remains
enabled.
WARNING
When a patient is left unattended, the bed should be lowered to its lowest position and the bed
elevation controls should be locked-out in order to reduce the risk of patient injury.
Содержание Spirit
Страница 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Страница 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Страница 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Страница 33: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 33 of 110 Section 3 Bed Operation ...
Страница 81: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 81 of 110 Section 4 Care Maintenance ...
Страница 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...