10.
Leaving the calibration masses in the bed center, place the bed in the maximum Trendelenburg position.
Press and hold the TREND button. Press ENTER button once the bed automatically stops in the maximum
Trendelenburg position. A confirmation “chirp” will sound to indicate that this step was successful.
11.
Leaving the calibration masses in the bed center, place the bed in the maximum reverse Trendelenburg
position. Press and hold the rev.TREND button. Press ENTER button once the bed automatically stops in
the maximum reverse Trendelenburg position. A confirmation “chirp” will sound to indicate that this step
was successful.
12.
After successful completion of the calibration procedure, a
message will be displayed. Press ENTER button to
continue. Press and hold the TREND button to level bed.
The bed will stop automatically when the mattress deck is
level with the floor. The scale has now been successfully
calibrated and the bed is now ready for use.
If the bed detects a problem during any stage of the scale
calibration procedure, a warning message will be displayed.
1.
Press ENTER button to acknowledge calibration
procedure failure.
2.
The display will automatically return you the
beginning of the calibration procedure.
3.
Press ENTER button to start calibration procedure
again or allow display to time out or remove and
replace footboard to abandon scale calibration
attempt at this time.
IMPORTANT: If the calibration procedure fails a second
time, the bed may require service and/or component
replacement.
WARNING
The Spirit™ bed may shift during Trendelenburg or Reverse Trendelenburg activation.
Initiate Trendelenburg or Reverse Trendelenburg only after the Central “Lock & Steer” system has
been put in the “BRAKE” position.
Содержание Spirit
Страница 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Страница 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Страница 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Страница 33: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 33 of 110 Section 3 Bed Operation ...
Страница 81: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 81 of 110 Section 4 Care Maintenance ...
Страница 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...