4.5
Troubleshooting Guide
NOTE:
Prior to using the Troubleshooting Guide, perform a brief function test on the bed.
Check all functions initiated from both the footboard staff control and the patient control
(staff/patient rail control and/or pendant) to determine which functions are not working.
Problem / Failure
Recommended Action
No functions working on bed
1.
Check to ensure bed is connected to AC power (plugged into wall
outlet)
2.
Check wall outlet for power
3.
Check Master Lockout. Hold all three patient lock-out buttons at
same time to lock/unlock, then try bed functions
4.
Lift head section of the mattress deck and remove dust shield
from on top of control box. Check for green power light (LED). If
no light, replace CB09 control box
5.
Call for Technical Assistance
Patient controls working
properly, footboard staff
control not working
1.
Check to ensure footboard is completely down into bracket
2.
Check cable connection under lower left side of footboard
3.
Check to make sure that there is no damage to Blindmate
connectors (plastic connector between bottom of footboard and
footboard mounting bracket)
4.
Check footboard staff control cable connection into DJB Box
under head section of mattress deck
5.
Replace footboard control
Footboard staff control
working properly, rail controls
not working
1.
Some functions may be locked out. Check patient lock-out
buttons on footboard. Patient lock-out prevents bed operation
from rail controls
2.
Check rail controls on both sides of bed. If only one rail control is
not working skip to next section. If both rail controls are not
working, continue to next step
3.
Check cable connections from rail to port on bed frame (ensure
jack is fully inserted into port)
4.
Check pendant connection into DJB Box under head section
5.
Replace DJB Box; Replace CB09 Control Box
Rail control working on one
side of the bed only
1.
Check cable connection (ensure jack is fully inserted into port)
2.
Unplug rail connection at port and replace with pendant (if
available). If pendant works, replace inoperable rail control. If
pendant doesn’t work, replace T-Cable
Footboard staff control
working properly, pendant
not working
1.
Some functions may be locked out. Check patient lock-out
buttons on footboard. Patient lock-out prevents bed operation
from the pendant
2.
Check cable connection from pendant to port on bed frame
3.
Check pendant connection into DJB Box under head section of
mattress deck
4.
Unplug pendant from current port and plug into port on opposite
side of bed. If pendant works in new port, replace DJB Box and T-
Содержание Spirit
Страница 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Страница 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Страница 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Страница 33: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 33 of 110 Section 3 Bed Operation ...
Страница 81: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 81 of 110 Section 4 Care Maintenance ...
Страница 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...