3.4.6
Nurse Call
The Spirit Plus™ and Spirit Select™ beds are equipped with a Nurse Call feature.
To operate the Nurse Call, press the red NURSE
button on the rail patient control (inside of each
head rail control).
Operation of the Nurse Call requires a conventional
connection cable (not supplied) to be plugged into
the ¼” Phono jack on the headboard mounting
bracket and into the designated outlet in the
hospital’s communication system.
The Spirit Plus™ beds are equipped with two Nurse
Call ports and Spirit Select™ beds are equipped
with one Nurse Call ports to accommodate multiple
Nurse Call controls.
Nurse Call ports are located on the headboard
mounting bracket.
Ensure that the attached “dummy plug” is inserted
to any inactive Nurse Call ports.