p/n 18603-001, Rev. J
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
Page 7-17
Symptoms
Possible Causes
What to Do
O
2
% is low.
O
2
inlet flow too low when
Low O
2
Source selected.
Verify the low pressure O
2
inlet has been
correctly calculated and set using the
Input O
2
Flow Chart
.
CareFusion recommends the use of an O
2
monitor to verify delivered O
2
%. Adjust the
entrained O
2
flow so the monitored value
shows the desired FIO
2
. (See the LTV
®
1200 Ventilator Operator's Manual for
information on using the Low O
2
Source
and O
2
% features.)
Rolled or leaking flow
valve inlet gasket.
Replace flow valve inlet gasket (see
Flow
Valve Assembly
in Chapter 8
).
Flow valve output is high.
Perform the Advanced FiO
2
Diagnostic
Procedures, see
page
7-27.
Failed calibration.
Recalibrate the vent. See
Chapter 6 -
Maintenance & Calibration
for instructions.
If the problem reoccurs after careful
recalibration, a transducer may be drifting
excessively. Replace the analog board.
See
Analog Board Assembly
in Chapter 8
for instructions.
Internal flexible tubing
occluded or pinched.
Internal flexible tubing has
pinhole leaks or leaking at
connections.
Open the vent and verify that none of the
flexible tubes connected to the solenoid
manifold, analog board, oxygen blender or
flow valve are pinched or leaking. See
Solenoid Manifold Assembly
or
Analog
Board Assembly
in Chapter 8 for a routing
diagram.
Defective or leaking
solenoid manifold.
Defective solenoid.
Replace the solenoid manifold. See
Solenoid Manifold Assembly
in Chapter 8
for instructions.
Incorrect O
2
blender
solenoid output flow.
Perform the Advanced FiO
2
Diagnostic
Procedures, see
page
7-27.
If testing fails, replace the O
2
blender, See
Chapter 8 - O2 Blender Assembly / O2
Block.
Содержание LTV 1100
Страница 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Страница 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 42: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page 6 3 Pneumatic Schematic Detail ...
Страница 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 315: ...Page C 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J This page left deliberately blank ...
Страница 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Страница 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 329: ...Page E 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map ...
Страница 330: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page E 3 Extended Features Map continued ...
Страница 331: ...Page E 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map continued ...
Страница 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...