FVd Delivered Volume:
1) Apply 15 cmH
2
O to the flow valve differential transducer at the high side port and verify
the
FVd
readout is within ± 0.5 cmH
2
O of the applied pressure.
2) Disconnect the tube from the syringe to the high pressure port of the valve differential
transducer so the connection is open to ambient air. When the displayed value
stabilizes, verify the
FVd
readout is within ± 0.5 cmH
2
O.
FDw Monitored Volume:
1) Apply 30 cmH
2
O to the Flow Differential transducer at the high side port and verify the
FDw
readout is within ± 0.5 cmH
2
O of the applied pressure.
2) Disconnect from the ventilator so the connection is open to ambient air. When the
display value stabilizes, verify the
FDw
readout is within ± 0.5 cmH
2
O.
FDb Inspired Volume
(LTM Only):
1) Apply 30 cmH
2
O to the Flow Differential transducer at the low side port and verify the
FDb
readout is within ± 0.5 cmH
2
O of the applied pressure.
2) Disconnect from the ventilator so the connection is open to ambient air. When the
display value stabilizes, verify the
FDb
readout is within ± 0.5 cmH
2
O.
O
2
Oxygen Inlet Pressure:
1) Apply 50 PSI to the high pressure oxygen inlet port and verify the readout is applied
pressure ± 1 PSI.
2) Disconnect from the ventilator so the connection is open to ambient air. When the
display value stabilizes, verify the readout is 0 ± 0.5 PSI.
Transducer Tolerance Variance Table:
Transducer
Acceptable
transducer
tolerance range
Recalibrate transducer if
variance to tolerance is;
Replace analog board
if variance to
tolerance is;
AP
± 0.5 cmH
2
O
>0.5 but < 2.5 cmH
2
O >2.5
cmH
2
O
FVd
± 0.5 cmH
2
O
>0.5 but < 0.75 cmH
2
O >0.75cmH
2
O
FDw
± 0.5 cmH
2
O
>0.5 but < 1.5 cmH
2
O >1.5
cmH
2
O
FDb
± 0.5 cmH
2
O
>0.5 but < 1.5 cmH
2
O >1.5
cmH
2
O
O
2
± 1.0 PSI
>1.0 but < 2.5 PSI
>2.5 PSI
If transducer recalibration is indicated, recalibrate that transducer (see
Chapter 6 -
Maintenance & Calibration
for instructions) and repeat the test.
If analog board replacement is indicated, replace the analog PCBA. See
Analog Board
Assembly
in Chapter 8 for instructions and repeat the test.
If the results for the transducers are within tolerance, perform the flow valve Accuracy
Checkout tests to determine accuracy of the flow valve.
Содержание LTV 1100
Страница 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Страница 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 42: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page 6 3 Pneumatic Schematic Detail ...
Страница 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 315: ...Page C 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J This page left deliberately blank ...
Страница 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Страница 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 329: ...Page E 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map ...
Страница 330: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page E 3 Extended Features Map continued ...
Страница 331: ...Page E 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map continued ...
Страница 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...