p/n 18603-001, Rev. J
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
Page 7-9
Symptoms
Possible Causes
What to Do
CONTINUED…
Monitored volume is
high.
Delivered volume is
high.
Defective flow valve.
Calibrate the flow valve. See
Flow Valve
Calibration
in Chapter 6 for instructions.
Clean the Flow Valve following the Flow
Valve Cleaning Procedure, see page 8-66.
Perform the Advanced Vte Diagnostic
Procedures, see
page 7-23
.
If testing and cleaning fails, replace the
Flow Valve.
See Chapter 8 - Flow Valve
Assembly and Cleaning Instructions.
Defective turbine.
Defective bypass valve.
Defective or leaking
solenoid manifold.
Perform the Advanced Vte Diagnostic
Procedures, see
page 7-23
.
If testing fails, replace defective parts.
Monitored volume is
low.
Delivered volume is
low.
Circuit leak.
Run a Leak Test and reseat or replace the
leaking parts or connections. See
Leak
Test
in Chapter 2 for instructions.
High or low side sense line
or elbow at exhalation
valve loose or leaking.
High or low sense lines are
occluded.
High or low sense ports in
the wye are occluded.
Check high and low pressure sense lines
to be sure they are correctly attached and
securely seated at both the ventilator and
wye ends.
Check the Luer fitting connections for
leaks.
Check the elbow connectors at the wye to
be sure they have not loosened or been
broken loose.
Verify lines are not occluded or pinched.
Exhalation drive line leaking
or loose.
Exhalation valve leaking
during inspiration.
Check the exhalation drive line at both the
ventilator and exhalation valve ends. Verify
the line is securely seated and not leaking.
Verify the exhalation valve is not leaking
during inspiration. If it is leaking, open the
exhalation valve and remove the
diaphragm. Reseat the diaphragm and
snap the valve cap back in place. See
Cleaning the Reusable Exhalation Valve
and Patient Circuit
in Chapter 4
for a
diagram of the exhalation valve assembly.
If necessary, replace the exhalation
diaphragm, or exhalation valve with a new
one.
Sense lines are reversed.
The sense lines are not designed to be
removed from either the wye or the Luer
fittings. If the sense lines have been
removed and replaced incorrectly, they
may not seal correctly when replaced.
Replace the patient wye and sense lines
with a known good assembly.
Содержание LTV 1100
Страница 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Страница 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 42: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page 6 3 Pneumatic Schematic Detail ...
Страница 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 315: ...Page C 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J This page left deliberately blank ...
Страница 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Страница 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 329: ...Page E 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map ...
Страница 330: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page E 3 Extended Features Map continued ...
Страница 331: ...Page E 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map continued ...
Страница 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...