p/n 18603-001, Rev. J
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
Page 7-37
Symptoms
Possible Causes
What to Do
DEFAULTS
alarm.
Event Log shows
DEFAULTS
.
Electro static discharge
(ESD)
.
Some or all control settings were found to
be invalid or out of range on power up and
were restored to the default settings.
Clear the alarm. Reduce static causing
conditions in the operating environment.
Defective main board.
Replace the main board. See
Main Board
Assembly
in Chapter 8 for instructions.
Defective memory board.
Replace the memory board. See
Memory
Board
in Chapter 8 for instructions.
Repeated
HIGH f
alarms
.
Total Breath Rate (
f
)
exceeds the set
HIGH f
alarm values.
Check
HIGH f
alarm values. (See the LTV
®
1200, 1150, or 1100 Ventilator Operator's
Manual for instructions).
Patient Circuit leak, causing
auto cycling.
Run a Leak test and reseat or replace the
leaking parts or connections. (See
Leak
Test
in Chapter 2 for instructions).
Internal problem with the
ventilator.
If problem reoccurs, immediately contact a
certified CareFusion service technician.
Repeated
HIGH PEEP
alarms.
Monitored PEEP exceeds
the set
HIGH PEEP
alarm.
Check
HIGH PEEP
alarm value. (See the
LTV
®
1200, 1150, or 1100 Ventilator
Operator's Manual).
Patient circuit and/or
Exhalation valve occluded.
Disassemble, clean and reassemble the
Patient Circuit and Exhalation Valve. See
Cleaning the Reusable Exhalation Valve
and Patient Circuit
in Chapter 4
for a
diagram of the exhalation valve assembly.
Internal problem with the
ventilator.
If problem reoccurs, immediately contact a
certified CareFusion service technician.
Remote Alarm
System does not work
with the ventilator.
Defective or improper
connections.
Check the Remote Alarm cable connection
between the ventilator’s Patient Assist Port
and the Remote Alarm System. (See the
LTV
®
1200, 1150, or 1100 Ventilator
Operator's Manual).
Defective Remote Alarm
cable.
Replace Remote Alarm cable. (See the
LTV
®
1200, 1150, or 1100 Ventilator
Operator's Manual for instructions).
Defective Remote Alarm
System.
Contact Remote Alarm System
manufacturer or service personnel.
Internal problem with the
ventilator.
Immediately contact a certified CareFusion
service technician.
Remote Alarm
System (
single
tone)
generates a pulsating
tone and
manufacturer’s
instructions indicate it
should be a
continuous tone.
PNT ASSIST option set to
PULSE
.
Set PNT ASSIST option to
NORMAL
. (See
the LTV
®
1200, 1150, or 1100 Ventilator
Operator's Manual).
Defective Remote Alarm
System.
Contact Remote Alarm System
manufacturer or service personnel.
Internal problem with the
ventilator.
Immediately contact a certified CareFusion
service technician.
Содержание LTV 1100
Страница 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Страница 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 42: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page 6 3 Pneumatic Schematic Detail ...
Страница 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 315: ...Page C 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J This page left deliberately blank ...
Страница 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Страница 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 329: ...Page E 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map ...
Страница 330: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page E 3 Extended Features Map continued ...
Страница 331: ...Page E 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map continued ...
Страница 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...