p/n 18603-001, Rev. J
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
Page 6-13
6) Connect the male Luer fitting from the calibration syringe assembly to the high pressure
female Luer fitting on the side of the ventilator. Decrease and maintain the vacuum until
a value in the range of 10 to 150 is displayed in the
Tidal Volume
window. If necessary,
add a large amount of compliance, such as a test lung, to the calibration syringe
assembly, this will significantly stabilize the readings during this step.
During this step, any value in the range of 10 to 150 displayed in the
Tidal Volume
window is equally acceptable, but it is very important that the value displayed in the
Tidal Volume
window is stable when pressing
Select
.
Alternatively, it is also acceptable to apply a small positive pressure to the ventilator low
pressure fitting to fulfill this step.
Record the number that was displayed when you pressed
Select
on the
Calibration Worksheet.
FD 0
C
M
H
2
O
is displayed.
7) Disconnect from the ventilator so the connection is open to ambient room air. Observe
the numbers displayed in the
Tidal Volume
setting window. When the displayed value is
stable, press
Select
.
Record the number that was displayed when you pressed
Select
on the Calibration
Worksheet.
FD 4015 ± 70 AD
is displayed.
8) Connect the male Luer fitting from the Calibration Syringe assembly to the high pressure
female Luer fitting. Increase and maintain the pressure until a value of 3945 to 4085 is
displayed in the
Tidal Volume
window. If necessary, add a large amount of compliance,
such as a test lung to the Calibration Syringe assembly, this will significantly stabilize the
readings during this step.
During this step, any value in the range of 3945 to 4085 displayed in the
Tidal Volume
window is equally acceptable, but it is very important that the value displayed in the
Tidal Volume
window is stable when pressing
Select
.
Record the number that was displayed when you pressed
Select
on the Calibration
Worksheet.
FD 0
C
M
H
2
O
is displayed
9) Disconnect from the ventilator so the connection is open to ambient room air. Observe
the numbers displayed in the
Tidal Volume
setting window. When the displayed value is
stable, press
Select
.
Record the number that was displayed when you pressed
Select
on the Calibration
Worksheet.
VALVE DIFF
is displayed if the calibration was successful.
If
CAL FAIL
appears at any time, press
Select
to exit the calibration. When
FLOW DIFF
is
displayed, press
Select
and perform the calibration again.
Содержание LTV 1100
Страница 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Страница 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 42: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page 6 3 Pneumatic Schematic Detail ...
Страница 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 315: ...Page C 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J This page left deliberately blank ...
Страница 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Страница 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 329: ...Page E 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map ...
Страница 330: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page E 3 Extended Features Map continued ...
Страница 331: ...Page E 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map continued ...
Страница 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...