![CareFusion LTV 1100 Скачать руководство пользователя страница 67](http://html1.mh-extra.com/html/carefusion/ltv-1100/ltv-1100_service-manual_2557142067.webp)
Page 6-28
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J
Servo
The Servo feature allows you to verify certain ventilator functions. When
SERVO
is on, the
ventilator continuously delivers gas according to the highlighted settings. Gas may be
delivered by:
turbine speed and flow valve step position
turbine speed and flow in Lpm
SERVO
is used to check ventilator operation or diagnose problems by setting specific
conditions on the ventilator and monitoring the system either externally (e.g. with a flow
meter or pressure gauge) or internally (by using the monitored values under the
RT XDCR
DATA
menu). For example, delivered flow can be verified
23
by setting the flow to a specified
Lpm
24
and connecting a flow meter to the patient wye. In addition, the monitored flow (
FTx
)
and other data can be checked under the
RT XDCR DATA
menu. The servo menu is set up
as follows:
SERVO
SERVO OFF
SERVO
ON
To enable the servo functions, press
Select
while
SERVO ON
is displayed. The following
displays are turned on (highlighted). Dimmed displays do not affect gas delivery.
Display Purpose
Tidal Volume
Sets the flow valve step position
Insp Time
Sets the flow in Lpm
O
2
%
Sets the delivered O
2
%
High Pres Limit / Low Pres
Sets the turbine speed
To turn off the servo functions, press
Select
while
SERVO OFF
is displayed. Servo
functions are automatically turned off when you enter the Calibration menu.
To modify a SERVO setting:
1) Press the associated button to select the control.
2) Turn the
Set Value
knob until the desired setting is displayed.
3) Press the button again or wait 5 seconds to accept the setting.
To select delivery by flow valve step position,
press
Tidal Volume
twice. The controls
for step position and turbine speed will be highlighted to indicate they are active.
To select delivery by Lpm,
press
Insp. Time
twice. The controls for flow, turbine speed
and O
2
% will be highlighted to indicate they are active.
Oxygen blending is active only when delivery by Lpm is selected.
To Home the flow valve,
press
Manual Breath
. After the home is complete, the valve will
return to its previous position.
23
For information on performance tolerances, see Appendix A - Ventilator Specifications in the LTV
®
1200 or the
LTV
®
1150 Ventilator Operator's Manual.
24
Turbine speed must be set to an appropriate value for flow to be delivered accurately.
Содержание LTV 1100
Страница 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Страница 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 42: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page 6 3 Pneumatic Schematic Detail ...
Страница 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 315: ...Page C 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J This page left deliberately blank ...
Страница 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Страница 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 329: ...Page E 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map ...
Страница 330: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page E 3 Extended Features Map continued ...
Страница 331: ...Page E 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map continued ...
Страница 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...