Page 9-24
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J
D) Turn the
PPin
solenoid off. Set
the turbine speed to 10 and
monitor the
PPP
transducer
reading.
Leakage rate must be
≤
1 cmH
2
O within
one minute
The
PPP
value under the
RT XDCR DATA
menu must be
0.5 cmH
2
O of the
measured pressure
E) Turn the
PPout
solenoid off.
Pressure at the exhalation drive port must
be 0.0
0.1 cmH
2
O.
F) Set
SERVO
to OFF. Apply 15
0.5 cmH
2
O to the high flow
XDCR port, and turn the
Apres
solenoid on. After reading, turn
the
Apres
solenoid off.
AP
and
FDw
display must read 0
0.5
cmH
2
O after activating the solenoid.
G) Apply and maintain 15
0.5
cmH
2
O pressure to the low and
high flow XDCR port, turn the
Purge
solenoid on for 3 sec, and
then turn it off.
FDw
display must read 0 ± 0.5 cmH
2
O
before and after activating the solenoid.
H) Apply 15
0.5 cmH
2
O to both
the low and high flow XDCR
ports, and turn the
ExhDiffP
solenoid on. After reading, turn
the
ExhDiffP
solenoid off.
FDw
display must read 0
0.5 cmH
2
O
before activating the solenoid and must
equal
AP
0.5 cmH
2
O after.
I) Apply and maintain 15 cmH
2
O
pressure to the high flow XDCR
port.
FDw
display must read 15
0.5 cmH
2
O.
10) Perform a Watchdog test
89
.
A) Unit under test:
AC adapter
Patient circuit
attached
Test lung(s)
attached
B) Navigate to
WDOG TEST
under
the
VENT MTNCE
menu and
press
Select
.
Ventilator powers down, powers up,
and successfully performs POST
Ventilator alarms, constant audible
alarm and flashes
RESET
alarm
Clear
alarms
89
It is normal for a
RESET
alarm to occur at the conclusion of POST after performing the Watchdog test, Battery
Duration test, or any other test that which causes the ventilator to go inoperative (other than pressing and
holding the On/Standby button). Press the Silence/Reset button twice to clear the alarm.
Содержание LTV 1100
Страница 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Страница 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 42: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page 6 3 Pneumatic Schematic Detail ...
Страница 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 315: ...Page C 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J This page left deliberately blank ...
Страница 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Страница 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 329: ...Page E 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map ...
Страница 330: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page E 3 Extended Features Map continued ...
Страница 331: ...Page E 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map continued ...
Страница 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...