Page 9-14
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J
2) Perform a Battery Duration/Battery Charge test
83
.
A) Operate unit from its internal battery
until it shuts down. Confirm
Vent
Inop
LED is lit and audible alarm is
sounding. Connect external power,
clear alarms and allow the unit to
operate normally and charge the
battery for a minimum of 8 hours.
BAT LOW, BAT EMPTY
, and
Vent Inop
alarms sound
When the unit shuts down confirm
Vent Inop
LED is lit and
Vent Inop
alarm sounds
B) With this charged battery, remove
external power, clear the
POWER
LOST
alarm and allow ventilator to
operate continuously, using the
following settings:
Mode:
Volume, Assist/Ctrl
Breath Rate:
15
Tidal Volume:
800
Insp. Time:
1.5 sec
PEEP:
5
Battery Level
LED is green upon
removing external power.
Unit must operate continuously for a
minimum of 40 minutes.
C) With ventilator ON, disconnect the
internal battery. Confirm
Vent Inop
LED is lit and alarm is sounding.
Reconnect external power (AC
adapter) and clear Reset alarm.
Charge Status
LED is red.
D) Reconnect the internal battery.
Charge Status
LED is flashing amber for a
few minutes (maximum 1 hour), and then
goes solid amber.
3) Perform an External Power test.
A) Continue with the external AC
adapter connected to the ventilator.
Ventilator runs normally.
External Power
LED shows green.
Charge Status
LED is lit or flashing.
B) Remove external power, clear
POWER LOST
alarm. Connect an
external DC power source with a
voltage between 14.8V and 15.0V,
using the external battery test cable.
Ventilator runs normally.
External Power
LED shows green.
Charge Status
LED is lit or flashing.
C) Slowly reduce the external DC
voltage until the
External Power
LED shows amber.
Ventilator runs normally.
External voltage is 11.0V
0.220 (2%)
Charge Status
LED is lit or flashing.
A
POWER LOW
alarm occurs.
Audible alarm sounds.
D) Slowly reduce the external DC
voltage until the
External Power
LED extinguishes.
Ventilator runs normally.
External voltage is 9.5V
0.190 (2%)
Charge Status
LED is off.
A
POWER LOST
alarm occurs.
Audible alarm sounds.
Battery Level
LED is lit.
83
It is normal for a
RESET
alarm to occur at the conclusion of POST after performing the Watchdog test, Battery
Duration test, or any other test that which causes the ventilator to go inoperative (other than pressing and
holding the On/Standby button). Press the Silence/Reset button twice to clear the alarm.
Содержание LTV 1100
Страница 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Страница 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 42: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page 6 3 Pneumatic Schematic Detail ...
Страница 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 315: ...Page C 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J This page left deliberately blank ...
Страница 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Страница 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 329: ...Page E 2 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map ...
Страница 330: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page E 3 Extended Features Map continued ...
Страница 331: ...Page E 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J Extended Features Map continued ...
Страница 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Страница 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...