CAPINTEC, INC.
CAPTUS
®
3000
2. Enter a descriptive name for the protocol being defined in the
Protocol Name
text
box. (32 characters max)
3. Input the number of measurements that will be made for the protocol in the
Number
of Measurements
text box. (1 to 50)
4. Click the appropriate radio button in the
Count Method
frame to select a count
method (Live time counting or Real time counting).
5. Click the appropriate check box(es) in the
Required Fields
frame to select the
demographic fields that will be required. Fields that are not checked will be optional
entries when adding or editing a patient.
6. Click the appropriate check box(es) in the
Optional Fields
frame to select any of the
optional fields, which will be needed for the protocol. Optional fields, which are
checked, will appear when adding or editing a patient and will be required entries.
Optional fields, which are not checked, will not appear on the Patient entry form.
7. If your protocol needs values, which are not displayed, click the
button. The screen will appear as shown in Figure 16-4.
Figure 16-4
8. Up to 5 additional fields may be entered. These are all numeric fields. Enter a name
(32 characters max) for each field that will be needed. Fields that are named will be
available when a patient is entered. If a name is not entered for a field, it will not be
displayed when a patient is entered.
9. Click the
button. The screen will appear as shown in Figure 16-5.
16-4
CUSTOM PROTOCOLS
November 13
Содержание CAPTUS 3000
Страница 2: ......
Страница 10: ...CAPINTEC INC CAPTUS 3000 INDEX WARRANTY DECLARATION OF CONFORMITY TOC 8 TABLE OF CONTENTS July 14...
Страница 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 337: ......
Страница 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...