Figure 5-10
An unacceptable deviation is indicated in red as HIGH.
If Deviation is indicated as HIGH, check the following:
•
The positioning of the rod source, especially for the Probe detector.
•
Ensure that there are no other sources (including dosed patients) in the area.
•
Click the
button or press
Alt+T
at the bottom of the Constancy Test screen.
Calibrated Activity and Calibration Date should match your Cs137 standard.
If a problem is found with any of the three items above, then correct the problem and
repeat the Constancy Test.
If no problems are found with the above items, then the efficiency of the Cs137 standard
should be re-measured. If this does not correct the problem, contact Capintec’s only
Authorized Service Center for service.
16. Click the
button or press
Alt+E
to exit and return to Figure 5-1 Quality Assurance
QUICK TIPS:
•
To print current results for record keeping, click the
button or press
Alt+P
on Auto Calibration or Constancy Test screen after calibration is
completed.
•
To view or print results for all the Quality Assurance tests performed to date,
click the
button or press
Alt+V
(Refer to the
section on page 5-24).
July 14
QUALITY ASSURANCE
5-9
Содержание CAPTUS 3000
Страница 2: ......
Страница 10: ...CAPINTEC INC CAPTUS 3000 INDEX WARRANTY DECLARATION OF CONFORMITY TOC 8 TABLE OF CONTENTS July 14...
Страница 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 337: ......
Страница 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...