CAPINTEC, INC.
CAPTUS
®
3000
Pre- Dose Measurement
frame – This option allows patient urine to be counted before
administering radioactive Vitamin B-12. The urine sample collected from the patient
before administration is called the patient blank. It will be subtracted from the post-dose
urine sample to yield net patient counts. This is an important measurement for a patient
who has previously received a radioactive isotope that may be excreted in the urine. It
may be skipped during the measurement procedure.
Normal Range of Excretion
Frame –
Minimum
and
Maximum
– These numbers will
define a range of values for excretion that can be considered as medically normal. They
are entered as seen fit by the doctors. Out of range results indicate a medical condition
and will be flagged in the report. Entering these values is optional.
2. To input new values in any frame, click in the appropriate text box or radio button or use
the
Tab
key to move from one selection to another.
3. Click the
button to save changes or click the
button to discard the
changes and return to Figure 10-2 Schilling Test Main Screen.
QUICK TIPS:
•
The
Standard Dilution
,
Urine Sample Aliquot
and
Counting Time
fields
appear again while adding a patient for a Schilling Test procedure. Any
changes made on the Add Patient screen will override the default values
entered in the Setup.
•
For greater flexibility, set
Pre-Dose Measurement
to Yes and
Repetitions
to
2, as these additional steps may be skipped if desired, while performing the
measurements.
•
To return to previous screen without saving changes press the
ESC
key.
Adding Patients
Note: All fields in bold are required before you can begin the Schilling Test. However, you
may input the Patient Name, Date of Birth and Patient ID alone, click the
button and input the remaining information later. The Patient Status box on Figure
10-2 Schilling Test Main Screen will display the message “Data Partially Entered”.
1. Click the
icon on Figure 10-2 Schilling Test Main Screen. The Patient
Information screen appears as shown in Figure 10-4, with information to be entered in
three sections.
November 13
LAB TESTS
10-5
Содержание CAPTUS 3000
Страница 2: ......
Страница 10: ...CAPINTEC INC CAPTUS 3000 INDEX WARRANTY DECLARATION OF CONFORMITY TOC 8 TABLE OF CONTENTS July 14...
Страница 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 337: ......
Страница 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...