CAPINTEC, INC.
CAPTUS
®
3000
1. Click the
button or press
Alt+C
on Figure 5-1 Quality Assurance Main
Screen. Figure 5-11 Constancy Test Main Screen will appear.
Figure 5-11 Constancy Test Main Screen
2. If Cs137 efficiency has
never
been measured on your system, proceed to Step 3.
Otherwise, skip to Step 4.
3.
Measuring Cs137 Efficiency
– The Constancy Test requires Cs137 efficiency value as
measured on your system. This step enters the Cs137 efficiency value in the Isotope
Library, which is used for calculations every time Constancy Test is performed.
Note: This is a one-time procedure and should not to be repeated unless Constancy
Test is unsuccessful.
a. Click the
button at the top of the screen or press
Alt+C
. The screen
appears as shown in Figure 5-12. Calibrated Activity and Date are already entered
from QA Setup done above.
Note:
is enabled only after auto calibration has been performed that
day.
July 14
QUALITY ASSURANCE
5-11
Содержание CAPTUS 3000
Страница 2: ......
Страница 10: ...CAPINTEC INC CAPTUS 3000 INDEX WARRANTY DECLARATION OF CONFORMITY TOC 8 TABLE OF CONTENTS July 14...
Страница 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 337: ......
Страница 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...