CAPINTEC, INC.
CAPTUS
®
3000
AUTO CALIBRATION
Calibration is done with Cs137 and Eu152 standard sources.
1. Click the
button or press
Alt+A
on Figure 5-1 Quality Assurance Main
Screen. Figure 5-3 Auto Calibration Screen will appear. When the procedure is started,
the spectrum appears in the upper half of the screen as it is being collected and displays
the measurement time elapsed in the upper right corner. Note that the
button at the top of the screen is enabled only after calibration has been done at least
once on that day.
Figure 5-3 Auto Calibration Screen
2. Note that in the
Select Detector
frame,
PROBE
is selected by default. To change the
detector to Well, click the radio button marked
WELL
.
3. Click the
button or press
Alt+S
to start the calibration procedure. The
screen appears as shown in Figure 5-4.
Figure 5-4
July 14
QUALITY ASSURANCE
5-5
Содержание CAPTUS 3000
Страница 2: ......
Страница 10: ...CAPINTEC INC CAPTUS 3000 INDEX WARRANTY DECLARATION OF CONFORMITY TOC 8 TABLE OF CONTENTS July 14...
Страница 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 337: ......
Страница 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...