CAPINTEC, INC.
CAPTUS
®
3000
Figure 10-28
Default Values
frame
Administered to Standard Activity Ratio
– This is the standard dilution factor,
which is 1000 for this procedure.
Sample Volume
– This is the volume used for the samples counted in the well
counter, and should not exceed the useful counting dimensions of the well. A sample
volume of 4 ml is a good sample for the well detector provided with the CAPTUS
®
3000 system.
Note: The volume must be the same for both the patient’s sample and the standard
solution sample.
Counting Time
– This is the duration for counting, in seconds, each time a
measurement is performed. The default value is set to 60 seconds.
Repetitions
frame – This is the number of repetitions each time a measurement is made.
For two repetitions, the average value is used in calculations. The second repetition may
be skipped during the measurement procedure, if so desired.
10-20
LAB TESTS
November 13
Содержание CAPTUS 3000
Страница 2: ......
Страница 10: ...CAPINTEC INC CAPTUS 3000 INDEX WARRANTY DECLARATION OF CONFORMITY TOC 8 TABLE OF CONTENTS July 14...
Страница 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 337: ......
Страница 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...