CAPINTEC, INC.
CAPTUS
®
3000
3. Click the
button or press
Alt+M
. The screen appears as shown in
Figure 6-6 with the name of the isotope at the top.
Figure 6-6
4. Click in the text box or use the
Tab
key to move around. Input the
Calibration Activity
,
Calibration Date
and
Calibration Time
from the source label and
Counting Time
desired in seconds.
5. To change the detector from
PROBE
to
WELL
, click in the appropriate radio button in the
Select Detector
frame.
6. Click the
button or press
Alt+N
. The screen appears as shown in Figure 6-8.
Note: If calibration has not been performed for the current day for the selected detector,
you will be prompted with the screen as shown in Figure 6-7. If calibration is
performed now, the program will return to Figure 6-8 after calibration is complete.
Figure 6-7
6-6
ISOTOPES
November 13
Содержание CAPTUS 3000
Страница 2: ......
Страница 10: ...CAPINTEC INC CAPTUS 3000 INDEX WARRANTY DECLARATION OF CONFORMITY TOC 8 TABLE OF CONTENTS July 14...
Страница 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Страница 337: ......
Страница 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...