49
•
MANUAL “HLD to DRAIN” Cycle
: Sets the reprocessing cycle to transfer disinfectant from the
reservoir into the basin and immediately down the drain, all in one step. Note: The HLD CYCLE
COUNT will be rest to “0” when the
START
key is pressed. This allows the operator to easily replace
the high-level disinfectant, and automatically reset the cycle counter at the same time.
• Press the
MANUAL
“
HLD
to
DRAIN
” key, and the LED light will now illuminate.
• Press
START/STOP
key to begin transferring HLD disinfectant from the reservoir into the basin
and down the drain. The
START/STOP
light will illuminate green, the
HLD
to
DRAIN
key will
begin blinking, and the display screen will show
HLD
to
DRAIN
.
• The CER OPTIMA
®
system cover can be opened during the cycle; however, be careful of spraying
fluids. Wearing appropriate PPE is required.
• At completion of the cycle, press
START/STOP
to end. The light will now turn red, and the display
screen will show Cycle Completed. The printer will automatically generate a printed cycle report (if
auto-print enabled), or press
to manually generate a cycle report.
• Note: The cycle data is retained in the CER OPTIMA system until either the next cycle is started, or
until system power is shut off.
• The completed cycle data will be automatically transferred to the optional CER OPTIMA Data
Management Program (if connected) at the completion of the cycle.
•
MANUAL “RSVR to BASIN” Cycle
: Sets the reprocessing cycle to transfer disinfectant from the
reservoir into the basin.
• Press the
MANUAL
“
RSVR
to
BASIN
” key, and the LED light will now illuminate.
• Press
START/STOP
key to begin transferring HLD disinfectant from the reservoir into the basin.
The
START/STOP
light will illuminate green, the
RSVR
to
BASIN
key will begin blinking, and
the display screen will show
RSVR
to
BASIN
.
• The CER OPTIMA cover can be opened during the cycle; however, be careful of spraying fluids.
Wearing appropriate PPE is required.
• At completion of the cycle, press
START/STOP
to end. The light will now turn red, and the display
screen will show Cycle Completed. The printer will automatically generate a printed cycle report (if
auto-print enabled), or press
to manually generate a cycle report.
• Note: The cycle data is retained in the CER until either the next cycle is started, or until system power
is shut off.
• The completed cycle data will be automatically transferred to the optional CER OPTIMA Data
Management Program (if connected) at the completion of the cycle.