46
•
If detergent is preferred for the wash phase, then manually add the detergent when the “WASH” led light
is blinking and the display screen shows WASH.
• During the
HLD
phase, the basin disinfectant temp will be displayed in Celsius ±1°C.
• If the CER OPTIMA
®
system cover is opened during the reprocessing cycle, the cycle will pause and the
display screen will show
CLOSE COVER
. The CER OPTIMA system cover must be closed within two
minutes or the cycle will be aborted.
• If an error is detected during the cycle, the
ALARM
will sound and the
ALARM RESET
light will
blink continuously. The display screen will show the error; refer to the Troubleshooting section for
further details on resolving the error. Press the
ALARM RESET
key to clear the alarm. The CER
OPTIMA system will now automatically empty the basin of any fluids, purge all tubing lines and return
the unit so it is ready to run a new cycle. The LCD display screen will indicate when the unit is fully
cleared. Note: In an alarm condition, the reprocessing cycle will need to be repeated for the device.
• Upon cycle completion, the audible tone will sound, the
START/STOP
light will turn red, and the
display screen will show Cycle Completed. The printer will automatically generate a printed cycle
report (if auto-print enabled), or press PRINT to manually generate a cycle report.
• Note: The reprocessing cycle data is retained in the CER OPTIMA unit until either the next endoscope
reprocessing cycle is started, or until system power is shut off.
• The completed reprocessing cycle data will be automatically transferred to the optional CER OPTIMA
Data Management Program (if connected) at the completion of the cycle.
• Open the CER OPTIMA cover and confirm all tubings have remained connected to the endoscope
for the full duration of the endoscope reprocessing cycle. If a hookup tubing is disconnected from
the endoscope, then the endoscope reprocessing cycle must be repeated. The endoscope can only be
removed from the CER OPTIMA basin, if all of the hookup tubings have been confirmed to have been
attached during the reprocessing cycle.
Caution!
An endoscope having a loose or disconnected hookup tubing at the
completion of a reprocessing cycle must be treated as still being contaminated
and cannot be used for a patient procedure. The endoscope reprocessing cycle
must be repeated. Do not use this endoscope on a patient.
• Note: The
AUTOMATIC
“
FULL
” cycle parameters settings for
WASH, AIR
, and
ALCOHOL
are
retained in the CER OPTIMA’s memory. The next time the
AUTOMATIC
“
FULL
” cycle is selected,
the default settings will automatically be used.
•
AUTOMATIC “HLD/RINSE” Cycle:
Sets the reprocessing cycle to automatically include a rinse, HLD
exposure, post-rinses (HLD dependent) and air and alcohol endoscope flushing (if selected).
• Press the
AUTOMATIC
“
HLD/RINSE
” key, and the LED light will now illuminate.
• Press the
AIR
key multiple times to cycle through and select the desired air purge time of 0, 3, 5
and 10 minutes.
!