Use the supplied plug, to block off the unused connection point located on the disinfectant reservoir.
For all reprocessor models, plug the electrical cord FIRMLY into the power entry module on the right side of the
reprocessor. Plug the other end into a standard AC power outlet.
Caution!
Only outlets outfitted with a ground fault interrupter (GFI) can be used.
Power
Press the power switch to turn
ON
the power to the reprocessor.
Printer
The CER OPTIMA
®
Endoscope Reprocessing System uses thermal printer paper to document all machine
functions. Ensure the printer is loaded with a fresh roll of printer paper before operating the machine or performing
a cycle. The printer paper will show a small pink/red colored line when the paper roll is nearing empty. If the
paper is empty or runs out during a print cycle, the cycle performed can be reprinted by selecting the Print
touchpad. The cycle data is available to print until either the “
START
” touchpad is pressed which starts a new
reprocessing cycle, or until the CER OPTIMA unit is shut off. Refer to the “
Maintenance and Troubleshooting
”
section for details on changing the printer paper.
Alcohol
Only use 70% Ethyl Alcohol or Isopropyl Alcohol for end-of-cycle endoscope drying. The CER OPTIMA
system has a compartment that houses the alcohol bottle, which is located on the front left-hand side of the
CER OPTIMA unit and is labeled “
ALCOHOL
”. To open the alcohol compartment, press in on the small door
indentation and then release; the door will now pop open slightly. Open the door fully to reveal alcohol bottle.
The alcohol bottle should be checked daily as part of the Quality Assurance Test, and refilled or replaced if low.
The alcohol compartment is designed to allow the use of round or oblong sized 500ml volume alcohol bottles.
If the alcohol runs out during a reprocessing cycle, the alcohol sensor will detect a low flow condition and display
a “
LOW ALCOHOL
” message in the LCD display. The operator will now have the opportunity to replace or
refill the alcohol bottle. If the alcohol bottle is not addressed within 2 minutes, then the cycle will be aborted and
the endoscope cycle will need to be rerun. Refer to the “Maintenance and Troubleshooting” section for details on
replacing or refilling the alcohol bottle.
Basin Drain Screen
Make sure the basin drain screen is positioned over the CER OPTIMA basin drain opening (remove the label tape
that secured the drain screen in place during shipping). The basin drain screen is used to prevent any loose items
from entering the drain, which would restrict fluid flow and could damage the internal workings of the machine
resulting in the need for servicing.