MAINTENANCE & TROUBLESHOOTING
Introduction
If a problem occurs, please use this section to locate the cause. It is best to review main topics to help identify
the problem. Listed below are possible solutions to problems which might occur.
Note: Contact your local representative or Medivators Customer
Support (1-800-444-4729) if problem is not resolved after using
the Troubleshooting Guide.
System Maintenance
Alcohol
Only use 70% Ethyl Alcohol or Isopropyl Alcohol for end-of-cycle endoscope drying.
The CER OPTIMA
®
endoscope reprocessing system has a compartment that houses the alcohol bottle, which is
located on the front left-hand side of the CER OPTIMA unit and is labeled “
ALCOHOL
”.
The alcohol bottle should be checked daily as part of the Quality Assurance Test, and refilled or replaced if low.
The alcohol compartment is designed to allow the use of round or oblong sized 500ml volume alcohol bottles. If
the alcohol runs out during a reprocessing cycle, the alcohol sensor will detect a low flow condition and display
a ‘Low Alcohol” message in the LCD display. The operator will now have the opportunity to replace or refill
the alcohol bottle. If the alcohol bottle is not addressed within 5 minutes, then the cycle will be aborted and the
endoscope cycle will need to be rerun.
To replace or refill the alcohol bottle:
Step 1:
Locate a new bottle of 70% Ethyl Alcohol or Isopropyl Alcohol.
Step 2:
Open the Alcohol compartment by pressing in on
the small door indentation and then releasing; the
door will pop open slightly. Open the door fully
to reveal the alcohol bottle.
Step 3:
Carefully slide the alcohol bottle out of the alcohol
compartment, and then loosen and remove the alcohol bottle
cap and associated uptake tube.
CHAPTER 5
Figure 15