Caution!
The use of accessories, components and cables other than those
specified by Medivators, as replacement parts for internal components or for
external accessories and connections, may result in increased emissions or
decreased immunity of the CER OPTIMA
®
Endoscope Reprocessing System.
Caution!
The pins of any connectors identified with an ESD warning symbol
should not be touched.
Connections, with this symbol, should not be made unless ESD precautionary
procedures are used.
Note: All personnel involved in the operation, installation and or maintenance of
the CER OPTIMA Endoscope Reprocessing System received and explanation of the
ESD warning symbol and training in ESD precautionary procedures.
Endoscope Hookups
Endoscopes attach to the CER OPTIMA basin connectors using specific tubing hookups. For proper endoscope
hook-up identification, refer to the interactive “Hookup Guide” located on the
Medivators website at
www.medivators.com. M
edivators Customer Support can also assist in endoscope hookup identification by
calling 1-800-444-4729.
EACH ENDOSCOPE HOOKUP IS INTENDED ONLY FOR CONNECTION WITH SPECIFIC ENDOSCOPE
MODELS IDENTIFIED IN THE MEDIVATORS ONLINE HOOKUP. PROPER ENDOSCOPE MODEL AND
HOOKUP COMBINATION, INCLUDING CONNECTIONS FOR ALL APPROPRIATE ENDOSCOPE CHANNELS
THAT REQUIRE A SEPARATE REPROCESSING CONNECTION, MUST BE VERIFIED PRIOR TO CONNECTION
.
The online hookup guide is available at: http://www.medivators.com/HookupLookup/.
Refer to the online hookup guide for additional notes and considerations specific to individual endoscope
models.
• Endoscopes must be thoroughly cleaned according to the manufacturer’s instructions and established
professional society guidelines prior to hookup connection. Failure to properly clean and prepare
endoscopes may result in inadequate high level disinfection and/or damage to the endoscope.
!
!