47
• Press the
ALCOHOL
key multiple times to cycle through and select the desired alcohol flush
volume of:
•
CER-1 OPTIMA: 0, 30, and 60cc’s
•
CER-2 OPTIMA: 0, 30, 60 and 120cc’s
• Press
START/STOP
key to begin the
HLD/RINSE
cycle.
• If Endoscope ID, Operator ID, and/or Patient ID are activated, then they will automatically display on the
screen in sequence after
START
has been selected. Use the barcode reader to scan the endoscope, operator
and/or patient ID information. Press
ENTER
to bypass if no information is available. The barcode reader
allows the ID data to be recorded and printed, or transferred to the optional CER OPTIMA
®
Data Management
Program at the completion of the cycle.
• If MRC Verification is activated, then confirm the result of the
HLD MRC CHECK
by pressing
ENTER
for Yes, and the
ARROW
key for No. Refer to the test strip manufacturer’s directions for use for instructions
on testing the MRC of the HLD.
• The reprocessing cycle will now start, and the
START/STOP
light will illuminate green. The front panel
display will show disinfectant type, cycle time remaining in minutes (counting down as the cycle progresses),
and the reprocessing step in progress.
•
During the HLD phase, the basin disinfectant temp will be displayed in Celsius ±1°C.
• If the CER OPTIMA cover is opened during the reprocessing cycle, the cycle will pause and the display
screen will show
CLOSE COVER
. The CER OPTIMA cover must be closed within two minutes or the
cycle will be aborted and
CYCLE CANCELLED
will be displayed.
• If an error is detected during the cycle, the
ALARM
will sound and the
ALARM RESET
light will blink
continuously. The display screen will show the error; refer to the Troubleshooting section for further details
on resolving the error. Press the
ALARM RESET
key to clear the alarm. The CER OPTIMA will now
automatically empty the basin of any fluids, purge all tubing lines and return the unit so it is ready to run a
new cycle. The LCD display screen will indicate when the unit is fully cleared. Note: In an alarm condition,
the reprocessing cycle will need to be repeated for the device.
• Upon cycle completion, the audible tone will sound, the
START/STOP
light will turn red, and the display
screen will show Cycle Completed. The printer will automatically generate a printed cycle report (if auto-
print enabled), or press
to manually generate a cycle report.
• Note: The reprocessing cycle data is retained in the CER OPTIMA system until either the next endoscope
reprocessing cycle is started, or until system power is shut off.
• The completed reprocessing cycle data will be automatically transferred to the optional CER OPTIMA Data
Management Program (if connected) at the completion of the cycle.
• Open the CER OPTIMA unit cover and confirm all tubing have remained connected to the endoscope for the
full duration of the endoscope reprocessing cycle. If a hookup tubing is disconnected from the endoscope,
then the endoscope reprocessing cycle must be repeated. The endoscope can only be removed from the
CER OPTIMA basin, if all of the hookup tubings have been confirmed to have been attached during the
reprocessing cycle.