24
Figure 4. Rear of Model CER-2 OPTIMA Reprocessor
Water
See Water Supply Section
Drain
Connect the flexible drain hose to the fitting labeled “
DRAIN
” fitting on the back of the reprocessor
and attach the screw clamp.
The drain must be 12-inches below the drain fitting on the reprocessor.
Eliminate all loops in the drain hose in order for fluids to drain completely. An open drain or air gap is required.
Never use a dishwasher attachment for the CER OPTIMA
®
drain connection.
Incomplete draining can result in an alarm condition or dilution of HLD.
Electrical Cord
If using a High Level Disinfectant that requires heating, plug the electrical cord FIRMLY into the power entry
module located on the front of the disinfectant reservoir. Plug the other end into a standard 120 VAC AC power
outlet. Press the reservoir’s power switch to turn
ON
the power to the disinfectant reservoir’s heaters. Refer to
the Operation section on adjusting the disinfectant reservoir’s temperature.
Refer to Figure 8.
Connect the disinfectant reservoir tubing having the right-angled connector and the contained disinfectant filter,
to the connection point labeled “
CHEMICAL INLET
” located on the reprocessor’s back panel. Connect the
other end having the straight connector to any one of the mating disinfectant reservoir connectors.
Connect the disinfectant reservoir tubing having the right-angled connector and no disinfectant filter, to the
connection point labeled “
CHEMICAL RETURN
” located on the reprocessor’s back panel. Connect the other
end having the straight connector to any one of the mating disinfectant reservoir connectors.
Reservoir
tubing