36
•
AIR
: Shutting of all
AUTOMATIC
and
MANUAL
functions and then pressing the
AIR
button alone,
allows a separate
AIR
cycle to be performed. The
AIR
time is pre-selected by the operator from 0, 3, 5,
and 10 minutes, prior to starting the cycle.
•
ALCOHOL
: Shutting of all
AUTOMATIC
and
MANUAL
functions and then pressing the
ALCOHOL
button alone, allows a separate ALCOHOL cycle to be performed. The
ALCOHOL
volume is pre-
selected by the operator, prior to starting the cycle.
•
The CER-1 OPTIMA allows for 0, 30 or 60cc’s volume
•
The CER-2 OPTIMA allows for 0, 30, 60 or 120cc’s volume
•
COMBINATION of WASH, AIR and/or ALCOHOL
: Shutting of all
AUTOMATIC
and
MANUAL
functions and then pressing any combination of
WASH, AIR
and
ALCOHOL
buttons, allows a separate
cycle to be performed. The WASH, AIR and
ALCOHOL
settings are pre-selected by the operator, prior
to starting the cycle.
START/STOP
Pressing the
START/STOP
key, begins or ends the selected cycle.
• When a cycle in-progress is stopped, the CER OPTIMA
®
system will clear the basin of any residual
fluids, purge all tubing lines, and reset itself so it is ready for a new cycle. The LCD display screen will
indicate when the unit is clearing itself, and also once it is ready for a new cycle to be initiated.
Pressing the
key at the end of an
AUTOMATIC
or
MANUAL
cycle provides a hardcopy printout
summarizing the endoscope reprocessing cycle.
• If the
AUTO
print function is activated, then the test results will automatically print at the completion
of the cycle. Multiple printouts can be received if the
touchpad is pressed, until a new cycle
START
is initiated or until the unit is powered off.
• If the
MANUAL
print function is activated, then the operator will need to press
at the completion
of the reprocessing cycle. Additional printouts can be received if the
touchpad is pressed again,
until a new cycle
START
is initiated or until the unit is powered off.
• The printout documents all reprocessing cycle log data and pertinent test information including: CER
OPTIMA model and serial number, date & time, cycle performed, a blank line for each Endoscope ID,
Operator ID, and Patient ID (if activated), or the Endoscope ID, Operator ID, and Patient ID number if
using the barcode, HLD type, cycle settings, MRC verification (if activated), and all individual cycle
steps.
• The printer paper has a shelf life based upon its storage conditions. It is important to understand these
storage conditions to ensure a long printed cycle retention. Refer to “
Maintenance and Troubleshooting
”
chapter to review the printer paper storage conditions.