28
Place the rigid end of tube “A” into the disinfectant container and the opposite end to the inlet of the
transfer pump.
Connect tube “B” between the outlet of the pump and the reservoir.
Connect tube “B” between the outlet of the pump and the adapter.
• Turn the pump
ON
at the switch.
Note: Ensure that there are no leaks. If leaks are seen, turn the pump OFF.
Note: There must be 4 gallons (15.1L) of HLD in disinfectant reservoir
of Model CER-1 OPTIMA. There must be 5 gallons (18.9L) of HLD in the
disinfectant reservoir of Model CER-2 OPTIMA.
• Fill the disinfectant reservoir until all the disinfectant has been transferred and turn the pump
OFF
.
•
Flush the Transfer Pump with clear water. Refer to “
Flushing the Transfer Pump and all tubing
”.
Alternative Filling Methods
Method 1
•
For a CER-1 OPTIMA, pour 4 gallons (15.1L) of new high level disinfectant into the CER OPTIMA
®
basin, and initiate a Manual “
Basin to Reservoir
” cycle to transfer the disinfectant to the disinfectant
reservoir. Refer to “
Cycle Selection: Automatic or Manual
” section for further details on initiating the
Manual cycle.
•
For a CER-2 OPTIMA, pour 5 gallons (18.9L) of new high level disinfectant into the CER OPTIMA
basin, and initiate a Manual “
Basin to Reservoir
” cycle to transfer the disinfectant to the disinfectant
reservoir. Refer to “
Cycle Selection: Automatic or Manual
” section for further details on initiating the
Manual cycle.
Method 2
•
Pour 4 gallons (15.1L) of disinfectant into the disinfectant reservoir of Model CER-1 OPTIMA.
•
Pour 5 gallons (18.9L) of disinfectant into the disinfectant reservoir of Model CER-2 OPTIMA.