5-8
Confidential
8501-00-1800, Revision 05
Candela Corporation
AlexTriVantage Laser System
Situation/Symptom
Fault #
Reason
Solution
Fault 15 - Transmission
Fault
15.1
Transmission (Tx) is >110%.
Recalibrate laser after each
step in the order listed below
until a successful calibration is
achieved:
Clean the Handpiece.
Try another Handpiece (the
same size if available). If
this works, contact
Customer Service to replace
the bad part.
Replace the Delivery
System.
If fault message persists, call
Candela Technical Support.
Fault 18 - Circuit Cal
Fault
18.1
HD, CP CktCal Test Energy
incorrect or checksum
corrupt.
Contact Candela Technical
Support.
Fault 18 - Circuit Cal
Fault
18.2
DI Factor checksum corrupt.
Contact Candela Technical
Support.
Fault 21 - Code Update
Fault
21
Not currently implemented.
Fault 22 - Interprocessor
Comm
22
Not currently implemented.
Fault 23 - 1-Wire Fault
23
One Wire Communications
Failure.
Contact Candela Technical
Support.
Faults 25 - Pulse
Duration Fault
25
Pockels Cell didn’t control
pulse width correctly.
Table 5-2: Fault Message and Solutions for the AlexTriVantage Laser System
Содержание AlexTriVantage
Страница 1: ...Operator s Manual PN 8501 00 1800 Revision 05 May 2007 Candela Corporation AlexTriVantage 0123 ...
Страница 14: ...xiv Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 30: ...1 16 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 64: ...2 34 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 82: ...4 8 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 110: ...B 4 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 112: ...C 2 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...