8501-00-1800, Revision 05
Confidential
3-7
Candela Corporation
AlexTriVantage Laser System
7.
Put on safety eyewear.
8.
Select the Calibration button (see
Figure 3-7: Calibration Button
The system enters the Ready state. The Standby/Ready button turns green (see
Figure 3-8: System is in the Ready State
The Confirm Calibration Parameters dialog appears asking you to confirm the
parameters you have specified (see
).
Figure 3-9: Confirm Calibration Parameters Dialog
9.
Verify that the parameters selected are correct for the patient and select the
checkmark to confirm the settings. If the settings are not correct, select the X to
cancel.
10.
The system prompts you to press and hold the Footswitch (see
). This
action starts the calibration.