2-4
Confidential
8501-00-1800, Revision 05
Candela Corporation
AlexTriVantage Laser System
A microprocessor based system controller is used to monitor and direct all system functions.
Users of the laser select parameters and monitor operation via electronic controls and a
display panel. A computer terminal gives the service technician access to the system
controller, both to obtain information and to control system functions, for maintenance and for
troubleshooting.
The AlexTriVantage Laser system delivers laser energy at a wavelength of about 755 nm
with a nominal pulsewidth of 50 nanoseconds. The output of this laser is delivered to the area
of treatment by means of a lens coupled user replaceable optical fiber with a treatment
Handpiece attached to its distal end. A trigger switch (footswitch) is used to control the
delivery of pulses. The user may choose to deliver a single pulse each time the trigger switch
is depressed, or pulses may be delivered repetitively as long as the switch is depressed, at
repetition rates up to 5 pulses per second.
The AlexTriVantage Laser control panel enables the user to select the desired energy density
(fluence) level and repetition rate. The control panel is also used to enable or disable the
triggering of the laser, to initiate the calibration feature and to obtain feedback from the
system, such as the number of pulses delivered or spot size selected.
Under CDRH regulations, the AlexTriVantage Laser is classified as a Class IV laser. Under
the Medical Device Directive, the AlexTriVantage Laser is classified as a Class llb, non
invasive, active device. The electrical classification for the AlexTriVantage under EN 60601-
1, with regards to protection from electric shock is Class 1 with a Type B applied part. Under
EN60825-1, the AlexTriVantage Laser is classified as a Class 4 laser.
The laser system is equipped with interlocks that inhibit the lasing function if the remote
interlock circuit is opened or parts of the delivery system removed.
Содержание AlexTriVantage
Страница 1: ...Operator s Manual PN 8501 00 1800 Revision 05 May 2007 Candela Corporation AlexTriVantage 0123 ...
Страница 14: ...xiv Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 30: ...1 16 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 64: ...2 34 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 82: ...4 8 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 110: ...B 4 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 112: ...C 2 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...