3-10
Confidential
8501-00-1800, Revision 05
Candela Corporation
AlexTriVantage Laser System
Figure 3-15: The Laser is Being Used for Treatment
When the Footswitch is released, the Lasing symbol is replaced with the Ready state
in the Status Area.
17.
When the treatment is complete, select the Standby button to place the system in
Standby. The Standby/Ready button turns yellow (see
Figure 3-16: System is in Standby
18.
When the system is in Standby, remove the Distance Gauge from the Handpiece.
19.
Insert the Handpiece in the Calibration Port.
Note:
If the laser remains idle (unused) for more than 2 minutes while in Ready state,
the system automatically reverts to the Standby state.
Note:
If you choose to change the Fluence setting during the treatment by selecting the
up and down arrow buttons, a confirmation message displays before the new settings
take effect. Select the checkmark to confirm the new setting.
Notes:
To return the Pulse Counter to zero, press the Pulse Count Reset button for at
least two seconds.
The laser system will not allow treatment pulses until a calibration has been
performed after any one of the following conditions:
Laser is turned on.
Fluence or Pulse Duration parameters are changed.
Delivery System is changed.
The Handpiece was changed or became disconnected from the distal end of
the Delivery System cable.
Specific faults occur
The laser system is in Standby for more than 30 minutes.
Содержание AlexTriVantage
Страница 1: ...Operator s Manual PN 8501 00 1800 Revision 05 May 2007 Candela Corporation AlexTriVantage 0123 ...
Страница 14: ...xiv Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 30: ...1 16 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 64: ...2 34 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 82: ...4 8 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 110: ...B 4 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 112: ...C 2 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...