3-8
Confidential
8501-00-1800, Revision 05
Candela Corporation
AlexTriVantage Laser System
Figure 3-10: Press and Hold Footswitch Dialog
11.
Press and hold the footswitch to begin a system calibration.
When the calibration is complete, the Calibration Complete dialog displays.
12.
Release the Footswitch.
A dialog prompts you to remove the Handpiece from the Calibration Port.
13.
Remove the Handpiece from the Calibration Port.
A message appears in the System Status Message Area indicating the calibration
is complete.
The system reverts to Standby (the Standby/Ready turns yellow) when you
remove the Handpiece (see
Figure 3-11: Standby
14.
Insert the Distance Gauge onto the Handpiece (see
Note:
After a calibration is completed, the laser will remain in the Ready State.