![Candela AlexTriVantage Скачать руководство пользователя страница 89](http://html1.mh-extra.com/html/candela/alextrivantage/alextrivantage_operators-manual_3863568089.webp)
8501-00-1800, Revision 05
Confidential
5-7
Candela Corporation
AlexTriVantage Laser System
Situation/Symptom
Fault #
Reason
Solution
Fault 10 - Delivery
System Fault
10.2
No Handpiece detected while
in Ready.
Remove the Delivery
System Handpiece.
Reinstall the Delivery
System Handpiece and
calibrate the laser.
If fault message persists, call
Candela Technical Support.
Fault 10 - Delivery
System Fault
10.4
No Fiber detected while in
Ready.
Remove the Delivery
System and reinstall.
Calibrate the laser.
Replace the Delivery
System with a spare or new
Delivery System. Calibrate
the laser.
If fault message persists, call
Candela Technical Support.
Fault 12 - Energy out of
Range Fault
12.1
Head energy (HD) of last
treatment pulse was 14%
lower than expected head
energy (xHD).
Try to calibrate the laser
system.
Contact Candela Technical
Support.
Fault 12 - Energy Out of
Range Fault
12.2
Head energy (HD) of last
treatment pulse was 14%
higher than expected head
energy (xHD).
Try to calibrate the laser
system.
Contact Candela Technical
Support.
Fault 12 - Energy Out of
Range Fault
12.3
Head energy (HD) of last
treatment pulse > Max
Treatment HE energy of
800+mJ.
Try to calibrate the laser
system.
Contact Candela Technical
Support.
Fault 13 - Trigger Switch
Fault
13
The redundant trigger
switches were in two different
states for > 1 second while in
ready.
Contact Candela Technical
Support.
Fault 14 - Simmer Fault
14
The simmer did not start or
dropped to while in Ready.
Contact Candela Technical
Support.
Table 5-2: Fault Message and Solutions for the AlexTriVantage Laser System
Содержание AlexTriVantage
Страница 1: ...Operator s Manual PN 8501 00 1800 Revision 05 May 2007 Candela Corporation AlexTriVantage 0123 ...
Страница 14: ...xiv Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 30: ...1 16 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 64: ...2 34 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 82: ...4 8 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 110: ...B 4 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 112: ...C 2 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...