4-6
Confidential
8501-00-1800, Revision 05
Candela Corporation
AlexTriVantage Laser System
Calibration Procedure
Energy calibration is the automatically controlled procedure by which the laser calibrates the
energy output to deliver pulses at the energy density selected on the control panel. During
the calibration procedure, the Handpiece is inserted in the Calibration Port, the system is
pulsed, and the energy output of the delivery system is read by an internal laser energy
meter. The laser determines the high voltage level necessary to obtain each selectable
Energy setting, and stores this information. Depending on the status of the system, the
calibration can require as few as 2 pulses or as many as 20 pulses.
System Calibration Requirements
There are a number of system conditions that cause a calibration to be required. The system
will not allow treatment unless a calibration has been completed.
A calibration is required by the system when:
1.
The system is turned on.
2.
The user changed the control panel Energy field.
3.
The user changed the control panel Handpiece field.
4.
A warning or fault has occurred requiring a calibration before treatment can continue.
5.
The system has been in the Standby state for more than 30 minutes.
If a calibration is required, and the user attempts to treat by pressing the trigger switch, the
system will not allow the treatment pulses, and indicates a calibration is required by
displaying the following message:
INSERT HANDPIECE INTO CAL PORT
If this occurs, continue with the Calibration Procedure below.
Initiating a Calibration
Sometimes it may be necessary to initiate a calibration. If the user desires a calibration OR if
the Handpiece was removed and replaced with another Handpiece having the same spot
size, a calibration should be initiated.
A calibration is initiated by changing the Energy field one setting lower (or higher), and then
putting the Energy field back to its original setting. Then, complete the Calibration Procedure
as normal.
Calibration Procedure
1.
Be sure that the desired Handpiece is attached to the optical fiber. Select the
Handpiece on the control panel. (The control panel handpiece selection and the
Handpiece must match for proper calibration.)
Содержание AlexTriVantage
Страница 1: ...Operator s Manual PN 8501 00 1800 Revision 05 May 2007 Candela Corporation AlexTriVantage 0123 ...
Страница 14: ...xiv Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 30: ...1 16 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 64: ...2 34 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 82: ...4 8 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 110: ...B 4 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 112: ...C 2 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...