8501-00-1800, Revision 05
Confidential
1-15
Candela Corporation
AlexTriVantage Laser System
Delivery System Cable Indicator
There is an Indicator on the Distal end of the Delivery System Cable that illuminates when the
laser is in Ready State. This Indicator illuminates at the same time as the Ready Indicator on
the front panel.
For more information, see
“Delivery System Cable” on page 2-13
Standby and Ready Operating Modes
The system operates in one of two states: Standby and Ready.
In the Standby state, laser emission is disabled. You must put the system into the Ready
state in order to enable laser emission.
In the Ready state, laser pulses are generated by depressing the Footswitch. As a safety
precaution, there is a delay of two seconds from the time that the system enters the Ready
state to the time that the laser emission is enabled.
When the laser system is not being used, it should be returned to the Standby state. The
laser will automatically revert back to the Standby state after 2 minutes of inactivity in the
Ready state.
You select the operating state via the Touch Screen/Display Panel. System state information
is displayed on the Status Area of the Touch Screen/Display Panel. When the system is in
the Ready state, the Ready Indicator on the front panel and the Delivery System Cable
Indicator are also illuminated.
For more information, see
“Calibration Button” on page 2-23
and
Button and Status Area” on page 2-23
.
Remote Interlock
An external connector for a remote interlock switch is provided on the rear panel of the
system. It is located on the lower left side of the rear panel. This interlock switch can be
connected to the doors of the laser room. If the door is opened while the laser is in the
READY state, the laser will return immediately to the STANDBY state where the laser beam
is extinguished.
For more information, on installation of a remote interlock, call Candela Technical
“Remote Interlock” on page 2-11
Note:
A Software Lock Button is included to prevent unauthorized use when not in
room. The laser will remain in warm-up mode when the software lock is engaged. The
software lock is user initiated.
Содержание AlexTriVantage
Страница 1: ...Operator s Manual PN 8501 00 1800 Revision 05 May 2007 Candela Corporation AlexTriVantage 0123 ...
Страница 14: ...xiv Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 30: ...1 16 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 64: ...2 34 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 82: ...4 8 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 110: ...B 4 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...
Страница 112: ...C 2 Confidential 8501 00 1800 Revision 05 Candela Corporation AlexTriVantage Laser System ...