CRVW3 3-Channel Vibrating-Wire Datalogger
A-3
7.
After the OS is sent, a period of about 25 seconds is required for the
CRVW3 to fully load the OS and restart itself. Do not attempt to connect
to the device during this time.
8.
When the load is complete, connect to the device and confirm in the
Status
table with the
Table Monitor
that the proper OS was loaded.
There is a risk that communications may be lost from
LoggerNet
to the remote CRVW3 if something goes wrong
during
LoggerNet
OS load process. Ensure you have a solid
radio communications channel with the device before
sending the OS update using
LoggerNet
and
File Control
.
You can use
File Control
in
PC400
or
PC200W
in a similar way to that
described above to send an OS to the CRVW3.