CRVW3 3-Channel Vibrating-Wire Datalogger
99
station. Interpret the information shown in this screen the same way as
when using the
Troubleshoot
tab.
The
Off Line Analysis
tab is only available in
DevConfig
when
not
connected to a CRVW3.
Repeat the above process for channels 2 and 3 (individually) if applicable.
Rename the various files as needed to indicate the station and channel that they
represent. It is recommended that some
TimeSeries.bin
files be captured soon
after the installation of a CRVW3 station so that they can be kept as baseline
references that document how the sensor was operating at the time of
deployment.
12.1.3 Adjusting Device Settings
You can view and change the current settings on each CRVW3 device using
the
Connect Screen
. This is possible even when connections are made remotely
to the CRVW3 over radio links, including repeater links. Choose the
Settings
Editor…
option from the
Datalogger
menu. A
Settings
screen will appear that
works much like
DevConfig
. You can view settings for the device and change
them if necessary. Review the warning below before changing any settings.
You can also use
PakBusGraph
to view and edit changes to the device’s
settings. Highlight the device in the main PakBus screen and right-click on the
device, then choose
Edit Settings…
Make sure you have selected the correct
PakBusPort
in the
PakBusNetwork
dropdown (as defined in
Setup Screen
) to
view your CRVW3 devices.
Make sure you have collected all required data before making a change to the
settings. After you apply changes to a CRVW3’s device settings, the CRVW3
will reset itself to begin measuring with the new settings. Close the
Settings
screen since it is no longer connected to the CRVW3. You will need to allow
up to 20 seconds to elapse after pressing
Apply
before you can expect to
NOTE